The influence of boiling bubble on the acoustic field and bubble dynamics in histotripsy: A numerical investigation
- PMID: 40514098
- DOI: 10.1016/j.medengphy.2025.104367
The influence of boiling bubble on the acoustic field and bubble dynamics in histotripsy: A numerical investigation
Abstract
Boiling histotripsy (BH) is a non-invasive treatment technique that leverages cavitation effects to disintegrate soft tissue. However, the influence of boiling bubble on the acoustic field and bubble dynamics, which are key mechanisms underlying tissue destruction in BH, is not yet fully understood. This study aims to elucidate the acoustic field distribution and dynamics of a single bubble, demonstrate the generation of intrinsic cavitation bubbles, and predict the associated cavitation damage using a numerical model. The Westervelt equation was employed to simulate the nonlinear propagation of ultrasound pulses in biological tissues, considering varying boiling bubble sizes at the focus and initial acoustic pressures on the transducer surface. Moreover, the Keller-Miksis equation model, combined with the Voigt model, was used to simulate the nonlinear oscillation of bubbles. The presence of a boiling bubble results in significant acoustic reflection at the bubble interface, particularly a notably strong negative pressure. Simulation results show that intrinsic cavitation occurs in a bubble radius of 0.4 mm at an initial acoustic pressure of 2 MPa. Increasing the boiling bubble radius to 1.1 mm leads to the extension of cavitation bubbles toward the second layer of the standing wave field. Additionally, the initial acoustic pressure seems to influence intrinsic cavitation more significantly than the boiling bubble radius. Bubble dynamics, including its maximum bubble radii and expansion duration, is significantly affected by the rarefactional wave. This effect is attributed not only to the peak negative pressure but also to the waveform. In conclusion, the presence of a boiling bubble at the focus substantially modifies the acoustic field distribution, the formation of intrinsic cavitation clouds, and the characteristics of bubble dynamics. The boiling bubble radius and the initial acoustic pressure on the transducer surface are critical factors in modifying cavitation-associated tissue fractionation. Appropriate parameter settings for ultrasonic pulses may enhance the efficacy and reduce the potential complications in BH treatment.
Keywords: Acoustic field; Boiling histotripsy; Bubble dynamics; Intrinsic cavitation.
Copyright © 2025 IPEM. Published by Elsevier Ltd. All rights reserved.
Conflict of interest statement
Declaration of competing interest The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be constructed as a potential conflict of interest.
Similar articles
-
The interaction of shockwaves with a vapour bubble in boiling histotripsy: The shock scattering effect.Ultrason Sonochem. 2021 Jan;70:105312. doi: 10.1016/j.ultsonch.2020.105312. Epub 2020 Aug 18. Ultrason Sonochem. 2021. PMID: 32866882 Free PMC article.
-
Control of the dynamics of a boiling vapour bubble using pressure-modulated high intensity focused ultrasound without the shock scattering effect: A first proof-of-concept study.Ultrason Sonochem. 2021 Sep;77:105699. doi: 10.1016/j.ultsonch.2021.105699. Epub 2021 Jul 31. Ultrason Sonochem. 2021. PMID: 34371476 Free PMC article.
-
Bubble dynamics in boiling histotripsy.Ultrasound Med Biol. 2018 Dec;44(12):2673-2696. doi: 10.1016/j.ultrasmedbio.2018.07.025. Epub 2018 Sep 15. Ultrasound Med Biol. 2018. PMID: 30228043
-
For Whom the Bubble Grows: Physical Principles of Bubble Nucleation and Dynamics in Histotripsy Ultrasound Therapy.Ultrasound Med Biol. 2019 May;45(5):1056-1080. doi: 10.1016/j.ultrasmedbio.2018.10.035. Epub 2019 Mar 26. Ultrasound Med Biol. 2019. PMID: 30922619 Free PMC article. Review.
-
Histotripsy: an innovative approach for minimally invasive tumour and disease treatment.Ann Med Surg (Lond). 2024 Mar 5;86(4):2081-2087. doi: 10.1097/MS9.0000000000001897. eCollection 2024 Apr. Ann Med Surg (Lond). 2024. PMID: 38576932 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources