Utilizing image analysis by optical density to evaluate changes in hematoxylin and eosin staining quality after reagent overuse
- PMID: 40548479
- DOI: 10.1080/01478885.2025.2517906
Utilizing image analysis by optical density to evaluate changes in hematoxylin and eosin staining quality after reagent overuse
Abstract
Due to the prevalence of hematoxylin and eosin (H&E) staining in routine histological preparations, understanding the factors that impact stain color characteristics is vital to attain consistently high-quality stains. In the last decade, increased use of digital pathology and image analysis (specifically by optical density [OD] measures) has provided new ways of assessing staining precision. This paper combines data from two studies that tracked H&E staining quality in both nuclear and cytoplasmic components of 12 tissues (11 human and 1 porcine) by OD after overuse of H&E staining reagents from 5 vendors. Both studies showed a decrease in eosin stain intensity by OD and visual inspection (by a histologist) with reagent overuse. This trend varied in degree by tissue type and reagent vendor. Nonetheless, staining quality for H&E staining from all vendors and for all organs remained acceptable (but not always optimal) for microscopic evaluation by the College of American Pathologists (CAP) and National Society for Histotechnology (NSH) staining criteria when stained sections were reviewed by a board-certified veterinary pathologist.
Keywords: Hematoxylin and eosin; image analysis; optical density; quality control; reagent overuse; reproducibility; staining precision.