Abstract
Recently nebulised antibiotics (gentamicin and carbenicillin) have been used successfully to treat respiratory tract infection in patients with cystic fibrosis. No information exists, however, on the choice of nebuliser or the ideal mode of operation with antibiotic solutions, which are often viscous. The aerosol output, droplet size, and nebulisation time were assessed for four common brands of jet nebuliser (Bird, DeVilbiss, Inspiron, and Upmist) used to nebulise 2 ml (80 mg) and 4 ml (160 mg) of gentamicin solution (Garamycin, Kirby-Warrick) at four compressed gas flow rates (6, 8, 10, and 12 1 . min-1). There were considerable variations between the nebulisers, DeVilbiss and Upmist being most efficient in the release of respirable (less than 5 micron diameter) droplets. Droplet size and nebulisation time were inversely proportional to gas flow rate. Aerosol output and nebulisation time were increased by raising the volume fill from 2 to 4 ml, although nebulisation time could still be restricted to 12 minutes or less with DeVilbiss and Upmist at 12 1 . min-1. The output of drug in droplets of below 5 micron diameter ranged from 7.2 (SE 0.4) to 71.4 (4.3) mg, according to the type of nebuliser, flow rate, and volume fill. These studies suggest that for optimal drug delivery 4 ml gentamicin solution should be nebulised either at a fixed flow rate of 10-12 1 . min-1 or with a high flow compressor. Previous unsatisfactory clinical results with antibiotic aerosols may have been due in part to incorrect choice of nebuliser or inappropriate operating conditions, or both.
Similar articles
-
Evaluation of two commercial jet nebulisers and three compressors for the nebulisation of antibiotics.Arch Dis Child. 1994 Oct;71(4):335-8. doi: 10.1136/adc.71.4.335. Arch Dis Child. 1994. PMID: 7979528 Free PMC article.
-
Choice of nebulisers and compressors for delivery of carbenicillin aerosol.Eur J Respir Dis. 1986 Sep;69(3):160-8. Eur J Respir Dis. 1986. PMID: 3780896
-
Factors influencing the size distribution of aerosols from jet nebulisers.Thorax. 1983 Oct;38(10):755-9. doi: 10.1136/thx.38.10.755. Thorax. 1983. PMID: 6648854 Free PMC article.
-
Nebuliser therapy in critical care: The past, present and future.J Intensive Care Soc. 2024 Feb;25(1):78-88. doi: 10.1177/17511437231199899. Epub 2023 Sep 20. J Intensive Care Soc. 2024. PMID: 39323591 Free PMC article. Review.
-
Inhaled antibiotics: dry or wet?Eur Respir J. 2014 Nov;44(5):1308-18. doi: 10.1183/09031936.00090314. Epub 2014 Oct 16. Eur Respir J. 2014. PMID: 25323242 Review.
Cited by
-
Deposition of carbenicillin aerosols in cystic fibrosis: effects of nebuliser system and breathing pattern.Thorax. 1988 Apr;43(4):318-22. doi: 10.1136/thx.43.4.318. Thorax. 1988. PMID: 3406919 Free PMC article.
-
A radiometric study of factors affecting drug output of jet nebulizers.Indian J Pharm Sci. 2010 Jan;72(1):31-8. doi: 10.4103/0250-474X.62234. Indian J Pharm Sci. 2010. PMID: 20582187 Free PMC article.
-
Selecting and using nebuliser equipment.Thorax. 1997 Apr;52 Suppl 2(Suppl 2):S92-101. doi: 10.1136/thx.52.2008.s92. Thorax. 1997. PMID: 9155861 Free PMC article. No abstract available.
-
Evaluation of two commercial jet nebulisers and three compressors for the nebulisation of antibiotics.Arch Dis Child. 1994 Oct;71(4):335-8. doi: 10.1136/adc.71.4.335. Arch Dis Child. 1994. PMID: 7979528 Free PMC article.
-
Pulmonary drug delivery. Part II: the role of inhalant delivery devices and drug formulations in therapeutic effectiveness of aerosolized medications.Br J Clin Pharmacol. 2003 Dec;56(6):600-12. doi: 10.1046/j.1365-2125.2003.01893.x. Br J Clin Pharmacol. 2003. PMID: 14616419 Free PMC article. Review.