. 2025 Jul 3:11:20552076251351841.
doi: 10.1177/20552076251351841. eCollection 2025 Jan-Dec.

Accuracy of heart rate, pulse oxygen saturation, and blood pressure using a non-contact photoplethysmography-based mobile application: A potential tool for wellness monitoring

Affiliations

Accuracy of heart rate, pulse oxygen saturation, and blood pressure using a non-contact photoplethysmography-based mobile application: A potential tool for wellness monitoring

Gianvincenzo Zuccotti et al. Digit Health. .

Abstract

Background: Photoplethysmography (PPG) is an effective non-contact technique for measuring vital signs. This report aims to evaluate whether a non-contact PPG-based mobile application supports an appropriate level of accuracy in monitoring SpO2, HR, and BP, compared against clinically approved medical devices.

Methods: Adult volunteers (≥18 years, both male and female) were consecutively enrolled from September to November 2024. Remote photoplethysmography (rPPG) for data collection involves recording subjects' faces using the front camera of mobile devices for ∼1.5 minutes. Simultaneously, vital signs will be continuously collected for about 1.5 minutes using the reference devices alongside data collected via the comestai application.

Results: A total of 562 participants (279 F/283 M) were enrolled. The app demonstrated excellent accuracy for HR (MAE 2.96, MAPE 4.14, and accuracy 99.1%) and SpO2 (MAE 2.10, MAPE 2.12, and accuracy 93.4%; no participants with SpO2 < 95% were not present). Results for systolic blood pressure (SBP) (MAE 14.24, MAPE 11.33, and accuracy 61.3%) and diastolic blood pressure (DBP) (MAE 9.83, MAPE 11.66, and accuracy 56.0%) were more moderate. The ROC curve analysis confirmed the app's robust detection capabilities for HR (AUC 0.976) and SpO2 (AUC 0.947), while showing fair performance for SBP (AUC 0.668) and weaker results for DBP (AUC 0.578). Bland-Altman plots revealed mean differences of -0.83 for HR, 1.84 for SpO2, -4.49 for SBP, and 2.54 for DBP compared to conventional methods.

Conclusions: The non-contact PPG-based mobile app holds promise as an accessible, user-friendly tool for wellness monitoring. Solutions like comestai.app present both new opportunities and challenges in the wellness sector, emphasizing the need for continued research in this field.

Keywords: Non-contact photoplethysmography; mobile application; monitoring; vital signs; wellness.

PubMed Disclaimer

Conflict of interest statement

The author(s) declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: POA and SDP are co-founders of Come Stai S.p.A.

Figures

Figure 1.
Figure 1.
Video frame captured by comestai.app.
Figure 2.
Figure 2.
Bland-Altman plots of systolic and diastolic pressure (Panels a and b), heart rate (Panel c), and oxygen saturation (Panel c), measures made using the mobile application and the conventional method. For each one, the horizontal line at the center of the mean difference between the two methods is plotted and the other two horizontal lines represent the limits of agreement, calculated as the mean±1.96*standarddeviationofthedifferences that show the interval in which 95% of the differences between the two measurement methods fall.

References

    1. Bart R, Ishak WW, Ganjian S, et al. The assessment and measurement of wellness in the clinical medical setting: a systematic review. Innov Clin Neurosci 2018; 15: 14–23. PMID: 30588362; PMCID: PMC6292717. - PMC - PubMed
    1. Hunter J, Leeder S. Patient questionnaires for use in the integrative medicine primary care setting—a systematic literature review. Eur J Integr Med 2013; 5: 194–216.
    1. Nagykaldi Z, Voncken-Brewster V, Aspy Cet al. et al. Novel computerized health risk appraisal may improve longitudinal health and wellness in primary care: a pilot study. Appl Clinic Inform 2013; 4: 75–87. - PMC - PubMed
    1. Huygens MW, Swinkels IC, de Jong JD, et al. Self-monitoring of health data by patients with a chronic disease: does disease controllability matter? BMC Fam Pract 2017; 18: 40. Published 2017 Mar 20. - PMC - PubMed
    1. Amjad A, Kordel P, Fernandes G. A review on innovation in healthcare sector (telehealth) through artificial intelligence. Sustainability 2023; 15: 6655.