The effect of 3D printing technology combined with Morse taper and platform switching design on microleakage performance of dental implants
- PMID: 40659999
- DOI: 10.1007/s10266-025-01152-y
The effect of 3D printing technology combined with Morse taper and platform switching design on microleakage performance of dental implants
Abstract
This study aims to investigate the effects of a Morse taper connection combined with platform switching design and hydroxyapatite (HA) coating on the mechanical properties, bacterial leakage, bioactivity, and osteogenic properties of dental implants. Four groups of models (platform switching of 0 and 0.6 mm; Morse taper of 0° and 10°) were determined through three-dimensional finite element analysis and labeled as TI, P, T, and PT groups. The dental implants were fabricated using selective laser melting and then surface-modified with sandblasting, acid etching, and HA coating. The physicochemical properties of the implants were measured using scanning electron microscopy (SEM), X-ray diffraction (XRD), energy dispersive spectrometry (EDS), and profilometry. Bacterial leakage was detected by colony counting. Finally, the biocompatibility, bioactivity, and osteogenic capacity of the implants were assessed through cell culture, immunohistochemistry, and Western blot. Significant differences were observed between the PT and Ti groups in terms of morphology, mechanical properties, and bacterial leakage. HA coating significantly enhanced the bioactivity of the implants. Implants modified by sandblasting, acid etching, and HA coating exhibited better biocompatibility, bioactivity, and osteogenic capacity compared to unmodified implants. Additionally, the PT group showed a significant improvement in reducing bacterial leakage. The Morse taper connection combined with platform switching design places less stress on the implant system than conventional platform-docked implants. The narrowing of the abutment neck is more favorable for the attachment of osteoblast-associated cells.
Keywords: Microbial leakage; Morse taper; Osseointegration; Platform switching.
© 2025. The Author(s), under exclusive licence to The Society of The Nippon Dental University.
Conflict of interest statement
Declarations. Conflict of interest: The authors declare that there were no financial interests that are directly or indirectly related to the work submitted for publication.
Similar articles
-
Comparative Study on the Effect of Various Implant-Abutment Connections on Microbial Leakage.J Pharm Bioallied Sci. 2025 Jun;17(Suppl 2):S1307-S1309. doi: 10.4103/jpbs.jpbs_1928_24. Epub 2025 Jun 18. J Pharm Bioallied Sci. 2025. PMID: 40655641 Free PMC article.
-
Evaluation of the Effect of Various Surface Conditioning Methods and Attachment of Platelet-rich Fibrin on Dental Implant Surface: An In Vitro Study.J Contemp Dent Pract. 2025 Apr 1;26(4):373-376. doi: 10.5005/jp-journals-10024-3840. J Contemp Dent Pract. 2025. PMID: 40583426
-
Influence of abutment types on the success of two-unit implant-supported fixed dental prostheses: A three-dimensional finite element analysis and fatigue test combined with microleakage evaluation.J Prosthet Dent. 2025 Jun 27:S0022-3913(25)00521-9. doi: 10.1016/j.prosdent.2025.06.006. Online ahead of print. J Prosthet Dent. 2025. PMID: 40581521
-
Platform switching for marginal bone preservation around dental implants: a systematic review and meta-analysis.J Periodontol. 2010 Oct;81(10):1350-66. doi: 10.1902/jop.2010.100232. J Periodontol. 2010. PMID: 20575657
-
Performance of conical abutment (Morse Taper) connection implants: a systematic review.J Biomed Mater Res A. 2014 Feb;102(2):552-74. doi: 10.1002/jbm.a.34709. Epub 2013 May 9. J Biomed Mater Res A. 2014. PMID: 23533139
References
-
- Hämmerle CHF, Tarnow D. The etiology of hard- and soft-tissue deficiencies at dental implants: a narrative review. J Periodontol. 2018;89(Suppl 1):S291-s303. - PubMed
-
- Liu Y, Wang J. Influences of microgap and micromotion of implant–abutment interface on marginal bone loss around implant neck. Arch Oral Biol. 2017;83:153–60. - PubMed
-
- Dong J, Zhang Z, Zhou G. Effect of taper and inserting angle on stability of micro-implants in type II bone with three-dimensional finite element. Hua Xi Kou Qiang Yi Xue Za Zhi. 2014;32(1):13–7. - PubMed
-
- Kim S, et al. The impact of Morse taper implant design on microleakage at implant-healing abutment interface. Dent Mater J. 2022;41(5):767–73. - PubMed