An automated continuous flow procedure for the determination of enterokinase
- PMID: 40719
- DOI: 10.1016/0009-8981(79)90144-x
An automated continuous flow procedure for the determination of enterokinase
Abstract
A continuous flow method has been developed for the automatic determination of enterokinase in rat small intesstine mucosa and/or luminal content. Trypsinogen was first hydrolysed by enterokinase under conditions which minimize autocatalytic activation. L-benzoyl-arginine paranitroanilide was then added and split to paranitroaniline by the trypsin so formed. Liberated paranitroalinine was diazotized and converted by the Bratton-Marshall reagent (N-naphthyl ethylene diamine) to an azodye, with maximum absorption at 550 nm. This method of determination was found to be six times more sensitive than the direct p-nitroaniline determination method. 36 determinations can be made hourly.
Similar articles
-
The application of a new synthetic substrate to the determination of enteropeptidase in rat small intestine and human intestinal biopsies.Clin Chim Acta. 1980 Feb 14;101(1):69-76. doi: 10.1016/0009-8981(80)90057-1. Clin Chim Acta. 1980. PMID: 6987011
-
[Method of determining the enterokinase content in the contents of the small intestines].Lab Delo. 1979;(9):531-3. Lab Delo. 1979. PMID: 91707 Russian. No abstract available.
-
Pancreaticobiliary factors in the modulation of small intestinal enterokinase in the rat.Am J Physiol. 1986 Jan;250(1 Pt 1):G103-8. doi: 10.1152/ajpgi.1986.250.1.G103. Am J Physiol. 1986. PMID: 3510561
-
Enterokinase.Proc Soc Exp Biol Med. 1994 Jun;206(2):114-8. doi: 10.3181/00379727-206-43728. Proc Soc Exp Biol Med. 1994. PMID: 8208733 Review.
-