Large-screen image intensifier photofluorography compared with full-size screen-film technique in chest radiography
- PMID: 4072746
- DOI: 10.1177/028418518502600506
Large-screen image intensifier photofluorography compared with full-size screen-film technique in chest radiography
Abstract
Large-screen image intensifier photofluorography revealed a radiation dose and exposure time of about one third of those of a medium speed rare-earth screen/film (Trimax T4/XD) combination for full-size chest radiography. The resolving power of the former technique is somewhat poorer and there is a moderate electron-òptical distortion in the peripheral field. In clinical chest radiography photospot films showed excellent demonstration of the central areas of the image but slight deterioration in peripheral parts. The minification effect of the photofluorographic films is advantageous for visual perception of many anatomic details on chest radiographs. The type of viewing device for photospot films is important.
Similar articles
-
Comparison of image intensifier photofluorography and full-size radiography in abdominal angiography.Acta Radiol Diagn (Stockh). 1981;22(1):39-47. doi: 10.1177/028418518102200106. Acta Radiol Diagn (Stockh). 1981. PMID: 7257851
-
[Optimization of the thoracic radiograph. Experimental and clinical study on the use of various film/screen combinations using rare earths].Radiol Med. 1986 May;72(5):261-6. Radiol Med. 1986. PMID: 3715076 Italian.
-
Image-intensifier photofluorography and conventional chest radiography: comparison of diagnostic efficacy.AJR Am J Roentgenol. 1988 Mar;150(3):539-44. doi: 10.2214/ajr.150.3.539. AJR Am J Roentgenol. 1988. PMID: 3257605
-
[New analog and digital imaging techniques for chest diagnosis--principles--clinical value--economics].Aktuelle Radiol. 1993 Jan;3(1):6-13. Aktuelle Radiol. 1993. PMID: 8448232 Review. German.
-
[Can digital luminescent radiography substitute for conventional x-ray technics in chest radiography?].Rofo. 1990 May;152(5):501-9. doi: 10.1055/s-2008-1046912. Rofo. 1990. PMID: 2160680 Review. German.