Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2025 Aug 3;15(1):28296.
doi: 10.1038/s41598-025-12927-2.

Comparison of consistency between wireless and conventional wired monitoring systems in gastrointestinal endoscopy

Affiliations
Comparative Study

Comparison of consistency between wireless and conventional wired monitoring systems in gastrointestinal endoscopy

Peng Zhang et al. Sci Rep. .

Abstract

There is a lack of research on use and effectiveness of wearable monitoring devices in the setting of invasive procedures, such as gastrointestinal endoscopy. Our study aimed to compare the consistency between wireless and conventional wired monitoring systems in gastrointestinal endoscopy under sedation. One hundred and twenty patients undergoing gastrointestinal endoscopy with sedation/anesthesia were fitted with both wireless and conventional wired monitoring systems. Multiple physiological parameters, including heart rate (HR), pulse rate, peripheral oxygen saturation (SpO2) and systolic (SBP), diastolic (DBP) and mean blood pressure (MAP), were monitored and recorded simultaneously. The correlation coefficients were 0.999, 0.939, 0.606, 0.837, 0.934 and 0.941 for HR, pulse rate, SpO2, SBP, DBP and MAP, respectively (all p < 0.001). The HR exceeded its limits of agreements (LoAs) and clinically acceptable consistency thresholds in 56 (4.17%) and 41 (3.05%) data, pulse rate in 84 (6.0%) and 137 (9.80%) data, SpO2 in 116 (8.30%) and 18 (1.28%) data, SBP in 39 (6.32%) and 121 (19.61%)data, DBP in 34 (5.34%) and 52 (8.43%) data, and MAP in 28 (4.54%) and 57 (9.24%) data. The polar concordance rate at 30 degrees was 59.31% for SBP, 61.12% for DBP, and 68.55% for MAP according to polar plot trending analysis. This study demonstrated that the wireless monitoring system showed adequate agreement with wired system for HR and SpO₂, but suboptimal agreement for pulse rate, SBP, DBP, and MAP during gastrointestinal endoscopy. The blood pressure measurements meets the ISO 81060-2 accuracy criteria. Overall, wireless monitoring is generally acceptable for gastrointestinal endoscopy, but further validation is needed for high-risk patients.

Keywords: Bland-Altman plot; Consistency; Gastrointestinal endoscopy; Wired monitoring; Wireless monitoring.

PubMed Disclaimer

Conflict of interest statement

Declarations. Competing interests: The authors declare no competing interests.

Figures

Fig. 1
Fig. 1
Schematic diagram of the wireless monitoring system (CareClever). The illustration was created by the authors (P.Z. and J.Z.)
Fig. 2
Fig. 2
Study flowchart. (WMS = wireless monitoring system; CMS = conventional wired monitoring system).
Fig. 3
Fig. 3
Bland-Altman plots of heart rate (HR), pulse rate, peripheral oxygen saturation (SpO2), and systolic (SBP), diastolic (DBP) and mean blood pressure (MAP). (WMS = wireless monitoring system; CMS = conventional wired monitoring system).
Fig. 4
Fig. 4
Polar plot of systolic, diastolic and mean blood pressure (SBP, DBP and MAP). (WMS = wireless monitoring system; CMS = conventional wired monitoring system).

Similar articles

References

    1. Leenen, J. P. L. et al. Current evidence for continuous vital signs monitoring by wearable wireless devices in hospitalized adults: systematic review. J. Med. Internet Res.22, e18636. 10.2196/18636 (2020). - PMC - PubMed
    1. Churpek, M. M., Yuen, T. C. & Edelson, D. P. Predicting clinical deterioration in the hospital: the impact of outcome selection. Resuscitation84, 564–568. 10.1016/j.resuscitation.2012.09.024 (2013). - PMC - PubMed
    1. Bonner, O., Beardsall, K., Crilly, N. & Lasenby, J. There were more wires than him’: the potential for wireless patient monitoring in neonatal intensive care. BMJ Innov.3, 12–18. 10.1136/bmjinnov-2016-000145 (2017). - PMC - PubMed
    1. Senechal, E. et al. Wireless monitoring devices in hospitalized children: a scoping review. Eur. J. Pediatr.182, 1991–2003. 10.1007/s00431-023-04881-w (2023). - PMC - PubMed
    1. Duncan, H. P., Fule, B., Rice, I., Sitch, A. J. & Lowe, D. Wireless monitoring and real-time adaptive predictive indicator of deterioration. Sci. Rep.10, 11366. 10.1038/s41598-020-67835-4 (2020). - PMC - PubMed

Publication types

LinkOut - more resources