Flow-suppressed 2D spin-echo imaging with high tolerance to B1 inhomogeneity using hyperbolic secant pulses
- PMID: 40785529
- DOI: 10.1002/mrm.70032
Flow-suppressed 2D spin-echo imaging with high tolerance to B1 inhomogeneity using hyperbolic secant pulses
Abstract
Purpose: To demonstrate flow-suppressed two-dimensional (2D) spin-echo and spin-echo diffusion echo-planar imaging (EPI) sequences using hyperbolic secant (HS) pulses for both π/2 excitation and π refocusing.
Theory and methods: A theoretical framework to derive phase dispersion of moving spins under π/2 excitation and π refocusing using HS pulses was described. Numerical simulations were performed to verify the validity of the theoretical analysis. All experiments were performed on a 3T clinical scanner. Phantom and human-brain imaging was performed using 2D spin-echo sequence, and liver imaging was performed using 2D spin-echo diffusion EPI. The signal-to-noise ratio and residual blood flow signal of the proposed sequences were compared with those of conventional spin-echo sequences using sinc pulses.
Results: Results from human brain and liver images demonstrated that the proposed method substantially reduced blood flow artifacts. In the brain, venous blood flow was suppressed more effectively with the proposed method than with conventional spin-echo sequence using presaturation. In the liver, as compared with spin-echo sequence using sinc pulses, the proposed method showed noticeable attenuation of bright blood signals at low b-values, whereas the overall tissue signal in peripheral regions was higher. The signal-to-noise ratio was enhanced by 10% to 30%, indicating improved B1 tolerance due to the adiabatic π refocusing HS pulse.
Conclusion: Flow suppression and partial B1 insensitivity were achieved by replacing sinc pulses with HS pulses in conventional 2D spin-echo imaging and spin-echo diffusion EPI sequences. This approach may be particularly useful in various applications requiring reduced vascular signal contamination, such as liver and brain imaging.
Keywords: B1 insensitivity; DWI; flow suppression; hyperbolic secant pulse; spin‐echo MRI.
© 2025 The Author(s). Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine.
Similar articles
-
Single-shot multi-b-value (SSMb) diffusion-weighted MRI using spin echo and stimulated echoes with variable flip angles.NMR Biomed. 2024 Dec;37(12):e5261. doi: 10.1002/nbm.5261. Epub 2024 Sep 22. NMR Biomed. 2024. PMID: 39308034 Free PMC article.
-
Multidimensional RF pulse design using auto-differentiable spin-domain optimization and its application to reduced field-of-view imaging.Magn Reson Med. 2025 Jun 16:10.1002/mrm.30607. doi: 10.1002/mrm.30607. Online ahead of print. Magn Reson Med. 2025. PMID: 40523117
-
MarkVCID cerebral small vessel consortium: II. Neuroimaging protocols.Alzheimers Dement. 2021 Apr;17(4):716-725. doi: 10.1002/alz.12216. Epub 2021 Jan 21. Alzheimers Dement. 2021. PMID: 33480157 Free PMC article.
-
Magnetic resonance perfusion for differentiating low-grade from high-grade gliomas at first presentation.Cochrane Database Syst Rev. 2018 Jan 22;1(1):CD011551. doi: 10.1002/14651858.CD011551.pub2. Cochrane Database Syst Rev. 2018. PMID: 29357120 Free PMC article.
-
Systemic treatments for metastatic cutaneous melanoma.Cochrane Database Syst Rev. 2018 Feb 6;2(2):CD011123. doi: 10.1002/14651858.CD011123.pub2. Cochrane Database Syst Rev. 2018. PMID: 29405038 Free PMC article.
References
REFERENCES
-
- Li Z, Hu HH, Miller JH, et al. A spiral spin‐echo MR imaging technique for improved flow artifact suppression in T1‐weighted postcontrast brain imaging: a comparison with Cartesian turbo spin‐echo. AJNR Am J Neuroradiol. 2016;37:642‐647. doi:10.3174/ajnr.A4600
-
- Song JW. Flow‐related artifacts and pitfalls in magnetic resonance imaging/angiography in neuroradiology. In: Proceedings of the Annual Meeting of ISMRM, Singapore, 2016.
-
- Haacke E, Lenz G. Improving MR image quality in the presence of motion by using rephasing gradients. Am J Roentgenol. 1987;148:1251‐1258. doi:10.2214/ajr.148.6.1251
-
- Felmlee JP, Ehman RL. Spatial presaturation: a method for suppressing flow artifacts and improving depiction of vascular anatomy in MR imaging. Radiology. 1987;164:559‐564. doi:10.1148/radiology.164.2.3602402
-
- Ehman RL, Felmlee JP. Flow artifact reduction in MRI: a review of the roles of gradient moment nulling and spatial presaturation. Magn Reson Med. 1990;14:293‐307. doi:10.1002/mrm.1910140214
Grants and funding
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous