The perturbation of electron beam dose distributions at medium interfaces
- PMID: 4079868
- DOI: 10.1118/1.595660
The perturbation of electron beam dose distributions at medium interfaces
Erratum in
- Med Phys 1986 Nov-Dec;13(6):966
Abstract
The perturbation of electron beam dose distributions in the vicinity of medium interfaces is calculated. Two different solutions to the Boltzmann equation, one an energy averaged solution and the other a diffusion approximation solution, are presented for the calculation of the electron distribution in the vicinity of interfaces. An extension of the energy averaged formalism is presented for media with high atomic number. A new effective depth approximation is introduced to calculate the beam conditions beneath the interface. The transformation from electron distribution to dose distribution is made and dose calculations are compared with published electron beam dose measurements.
Similar articles
-
Dose distributions in regions containing beta sources: large spherical source regions in a homogeneous medium.Med Phys. 1988 May-Jun;15(3):358-63. doi: 10.1118/1.596231. Med Phys. 1988. PMID: 3405138
-
Optimization of pencil beam widths for electron-beam dose calculations.Med Phys. 1995 Apr;22(4):411-9. doi: 10.1118/1.597606. Med Phys. 1995. PMID: 7609721
-
Calculation of dose distributions in the vicinity of high-Z interfaces for photon beams.Med Phys. 1995 Oct;22(10):1685-90. doi: 10.1118/1.597433. Med Phys. 1995. PMID: 8551995
-
Dose perturbations at interfaces in photon beams.Med Phys. 1987 Jul-Aug;14(4):585-95. doi: 10.1118/1.596023. Med Phys. 1987. PMID: 3626998
-
Calculation of high-energy electron dose distributions in tissue-equivalent media. I. Determination of the dose function of point unidirectional sources.Strahlentherapie. 1982 May;158(5):298-304. Strahlentherapie. 1982. PMID: 7112625