Analysis of the Effect of Pupil Size and Corneal Aberration on the Optical Performance of Premium Intraocular Lenses
- PMID: 40806958
- PMCID: PMC12347176
- DOI: 10.3390/jcm14155336
Analysis of the Effect of Pupil Size and Corneal Aberration on the Optical Performance of Premium Intraocular Lenses
Abstract
Background/Objectives: To assess the optical performance of two refractive premium IOLs across pupil sizes and values of corneal spherical aberration (SA). Methods: Two refractive IOLs were evaluated in this study: Tecnis Eyhance and Mini Well. The surface profiles were obtained to calculate the through-object MTF (TO MTF) curves and simulate optotype images. Entrance pupil sizes ranging from 2 to 5.5 and three corneal models were analyzed in the simulation: an average population aberrated cornea, an aberration-free cornea and a post-Lasik myopic cornea. Results: For Model 1 and pupil sizes between 3.0 and 3.5 mm, Mini Well provided acceptable visual quality from far to near distances, whereas Eyhance struggled to maintain visual quality at distances closer than intermediate. For patients with lower-than-normal corneal SA (i.e., more prolate corneas, such as post-hyperopic LASIK) both IOLs exhibited a hyperopic shift in far focus. Conversely, for patients with higher-than-normal corneal SA (i.e., more oblate corneas, such as post-myopic LASIK), the shift occurred in the myopic direction. Despite the implementation of an optimized IOL power to circumvent any shift, the TO MTF nevertheless reflected the interaction between corneal and IOL SA. Furthermore, the Mini Well demonstrated increased tolerance to less negative SA values, while Eyhance exhibited behavior consistent with a monofocal lens for more positive SA values. Conclusions: Surgeons should consider each patient's corneal asphericity and typical pupil diameter when selecting and calculating the power of the premium IOLs studied, particularly in patients with a history of refractive surgery.
Keywords: Lasik surgery; cataract; premium IOLs; refractive surgery.
Conflict of interest statement
The authors declare no conflicts of interest.
Figures
References
-
- Standardization IOf. Ophthalmic Implants–Intraocular Lenses—Part 2: Optical Properties and Test Methods. International Organization for Standardization; Geneva, Switzerland: 2014.
-
- Tandogan T., Son H.S., Choi C.Y., Knorz M.C., Auffarth G.U., Khoramnia R. Laboratory Evaluation of the Influence of Decentration and Pupil Size on the Optical Performance of a Monofocal, Bifocal, and Trifocal Intraocular Lens. J. Refract. Surg. 2017;33:808–812. doi: 10.3928/1081597X-20171004-02. - DOI - PubMed
LinkOut - more resources
Full Text Sources
Miscellaneous