Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2025 Jul 21:10:100350.
doi: 10.1016/j.jhlto.2025.100350. eCollection 2025 Nov.

Alterations in implantable cardioverter defibrillator lead parameters following left ventricular assist device implantation

Affiliations

Alterations in implantable cardioverter defibrillator lead parameters following left ventricular assist device implantation

David Gittess et al. JHLT Open. .

Abstract

Background: Left ventricular assist devices (LVADs) are increasingly used in the management of advanced heart failure. The majority of these patients have pre-existing implantable cardioverter defibrillators (ICDs). The proximity between the LVAD inflow cannula and right ventricular (RV) defibrillation lead raises the potential for disruption of ICD function.

Methods: This is a retrospective analysis of 95 patients with ICDs at a single tertiary care center who underwent LVAD implantation and who met inclusion criteria. The primary outcome was changes in the pre-operative and post-operative transvenous ICD RV lead parameters. These changes were stratified by the age of the RV lead and analyzed via a paired t-test. The secondary outcome was disruption to the ICD requiring an intervention.

Results: LVAD implantation was associated with significant decreases in sensed amplitude (p < 0.01) and high voltage impedance (p < 0.01) and an increase in capture threshold (p = 0.017). When stratified by age of the RV lead, patients with leads older than two years had similar trends in all parameters. However, RV leads that were two years old or younger only showed a significant change in high voltage impedance (p < 0.01). Mechanical disruption of the ICD related to the surgery was infrequent but significant.

Conclusion: Because LVAD implantation is capable of impacting ICD function and causing mechanical disruption, close monitoring should be paid to the ICD in the peri-operative period including obtaining a full interrogation.

Keywords: Advanced heart failure; Electrophysiology; Implantable cardioverter defibrillator (ICD); Left ventricular assist device (LVAD); Right ventricular lead disfunction.

PubMed Disclaimer

Conflict of interest statement

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Figures

Figure 1
Figure 1
Flow diagram.

Similar articles

References

    1. Grimm M., Grimm G., Zuckermann A., et al. ICD therapy in survivors of sudden cardiac death awaiting heart transplantation. Ann Thorac Surg. 1995;59(4):916–920. doi: 10.1016/0003-4975(95)00013-b. - DOI - PubMed
    1. Moss A.J., Zareba W., Hall W.J., et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346(12):877–883. doi: 10.1056/NEJMoa013474. - DOI - PubMed
    1. Bardy G.H., Lee K.L., Mark D.B., et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005;352(3):225–237. doi: 10.1056/NEJMoa043399. - DOI - PubMed
    1. Agrawal S., Garg L., Nanda S., et al. The role of implantable cardioverter-defibrillators in patients with continuous flow left ventricular assist devices - A meta-analysis. Int J Cardiol. 2016;222:379–384. doi: 10.1016/j.ijcard.2016.07.257. - DOI - PubMed
    1. Clerkin K.J., Topkara V.K., Demmer R.T., et al. Implantable cardioverter-defibrillators in patients with a continuous-flow left ventricular assist device: an analysis of the INTERMACS registry. JACC Heart Fail. 2017;5(12):916–926. doi: 10.1016/j.jchf.2017.08.014. - DOI - PMC - PubMed

LinkOut - more resources