Comparative study by FEM of different liners of a transfemoral amputated lower limb
- PMID: 40883410
- PMCID: PMC12397369
- DOI: 10.1038/s41598-025-15974-x
Comparative study by FEM of different liners of a transfemoral amputated lower limb
Abstract
The mechanical behavior of prosthetic liners significantly influences stress distribution, soft tissue protection, and the overall efficiency of the prosthetic. While extensive research has been conducted on liner materials, the impact of liner thickness (2 mm, 4 mm, and 6 mm) on biomechanical response remains underexplored. This study utilizes finite element analysis in Abaqus to investigate how liner material (Gel vs. Silicone) and thickness affect contact pressure (CPRESS), maximum principal strain (Le. Max), shear stress (CSHEAR1), and vertical displacement (U3) at the residual limb-liner interface. A three-dimensional numerical model was developed to simulate stress transmission and displacement behavior under physiological loading conditions. The results demonstrate that liner thickness plays a critical role in modulating pressure distribution and mechanical stability, with Gel providing superior flexibility and shock absorption, whereas Silicone offers enhanced structural integrity. At a thickness of 2 mm, the highest pressure of 0.4656 MPa is recorded. When the thickness is increased to 4 mm, the pressure decreases to 0.4153 MPa, reflecting a reduction of approximately 10.8%. Further increasing the thickness to 6 mm results in a pressure drop of 0.3825 MPa, corresponding to a total reduction of 17.9%. These findings provide quantitative insights into stress attenuation mechanisms, contributing to the optimization of prosthetic liner design for improved clinical outcomes in lower-limb amputees.
Keywords: Contact pressure; Displacement; Finite element analysis; Prosthetic liner; Shear stress.
© 2025. The Author(s).
Conflict of interest statement
Declarations. Competing interests: The authors declare no competing interests.
Figures
References
-
- Boutwell, E., Stine, R., Hansen, A., Tucker, K. & Gard, S. Effect of prosthetic gel liner thickness on gait biomechanics and pressure distribution within the transtibial socket. J. Rehabil. Res. Dev.49(2), 227–240 (2012). - PubMed
-
- Bertolini, M., Moreschini, C., Siffredi, P., Colombo, G. & Rossoni, M. Finite element analysis of the donning phase of a prosthetic socket for transfemoral amputees. In Advances in Digital Human Modeling DHM 2023 Lecture Notes in Networks and Systems Vol. 744 (eds Scataglini, S. et al.) (Springer, 2023). 10.1007/978-3-031-37848-5_1.
-
- Ragolta, C.I. & Morford, M. Novel materials for prosthetic liners. NASA Kennedy Space Center. https://ntrs.nasa.gov/api/citations/20110016246/downloads/20110016246.pdf (2011).
-
- Lagan, S. & Liber-Kneć, A. The determination of mechanical properties of prosthetic liners through experimental and constitutive modelling approaches. Tech. Trans.3, 197–209 (2018).
-
- Ali, I., Kumar, R. & Singh, Y. Finite element modelling and analysis of trans-tibial prosthetic socket. Glob. J. Res. Eng.14, 42–50 (2014).
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Medical