Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2025 Dec 2;25(23):7338.
doi: 10.3390/s25237338.

Assessment of Image Quality Performance of a Photon-Counting Computed Tomography Scanner Approved for Whole-Body Clinical Applications

Affiliations

Assessment of Image Quality Performance of a Photon-Counting Computed Tomography Scanner Approved for Whole-Body Clinical Applications

Francesca Saveria Maddaloni et al. Sensors (Basel). .

Abstract

Background: Photon-counting computed tomography (PCCT) represents a major technological advance in clinical CT imaging, offering superior spatial resolution, enhanced material discrimination, and potential radiation dose reduction compared to conventional energy-integrating detector systems. As the first clinically approved PCCT scanner becomes available, establishing a comprehensive characterization of its image quality is essential to understand its performance and clinical impact.

Methods: Image quality was evaluated using a commercial quality assurance phantom with acquisition protocols typically used for three anatomical regions-head, abdomen/thorax, and inner ear-representing diverse clinical scenarios. Each region was scanned using both ultra-high-resolution (UHR, 120 × 0.2 mm slices) and conventional (144 × 0.4 mm slices) protocols. Conventional metrics, including signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), slice thickness accuracy, and uniformity, were assessed following international standards. Task-based analysis was also performed through target transfer function (TTF), noise power spectrum (NPS), and detectability index (d') to evaluate diagnostic relevance.

Results: UHR protocols provided markedly improved spatial resolution, particularly in the inner ear imaging, as confirmed by TTF analysis, though with increased noise and reduced low-contrast detectability in certain conditions. CT numbers showed linear correspondence with known attenuation coefficients across all protocols.

Conclusions: This study establishes a detailed technical characterization of the first clinical PCCT scanner, demonstrating significant improvements in terms of spatial resolution and accuracy of the quantitative image analysis, while highlighting the need for noise-contrast optimization in high-resolution imaging.

Keywords: CT image quality; photon-counting detectors; task-based metrics.

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflicts of interest.

Figures

Figure A1
Figure A1
Low-contrast module of the Catphan600 phantom shown for some of the acquired districts: (a) Head; (b) Abdomen; (c) Inner Ear at standard mode (144 × 0.4 mm); (d) Inner Ear at UHR (120 × 0.2 mm). Window/Level settings were chosen to achieve the best visibility possible of the low-contrast objects.
Figure A2
Figure A2
CTP404 module shown for the Inner Ear acquisitions: (a) standard (144 × 0.4 mm), (b) UHR (120 × 0.2 mm) acquisitions. The red arrow shows the polystyrene insert used for TTF calculation.
Figure 1
Figure 1
(a) 128 × 128 noise ROI for head protocol; (b) noise ROIs for all the analyzed thorax/abdominal protocols; (c) noise ROIs for all the analyzed inner ear protocols.
Figure 2
Figure 2
(a) HU distribution for head protocol; (b) HU distribution for abdomen protocol; (c) HU distribution for inner ear protocol.
Figure 3
Figure 3
Representation of NPS2D for (a) head, (b) abdomen, and (c) inner ear UHR. Axis ticks refer to the spatial frequencies in x–y directions.
Figure 4
Figure 4
(a) Comparison between standard and flash abdomen protocols; (b) NPS curves for two acquisition collimation widths (144 × 0.4 mm and 120 × 0.2 mm) in the inner ear protocol; (c) comparison between two reconstruction kernels (Hr68 and Hr72) at two different CTDIvol values for inner ear; (d) NPS for the head protocol.
Figure 5
Figure 5
(a) TTF curves for inner ear at 140 kV, 200 mAs comparing 144 × 0.4 mm and 120 × 0.2 mm collimation widths; (b) TTF curves for inner ear at 100 kV, 300 mAs comparing 144 × 0.4 mm and 120 × 0.2 mm collimation widths; (c) TTF curves for thorax/abdominal protocol; (d) TTF curves for thorax/abdominal protocol comparing standard and flash acquisitions; (e) TTF curves for the head protocol.
Figure 6
Figure 6
Linearity of the HUs vs. linear attenuation coefficients computed for the specific spectrum, for abdominal (a) and head (b) protocols.

References

    1. Brenner D.J. Computed Tomography—An Increasing Source of Radiation Exposure. N. Engl. J. Med. 2007;357:2277–2284. doi: 10.1056/NEJMra072149. - DOI - PubMed
    1. Booij R., Budde R.P.J., Dijkshoorn M.L., van Straten M. Technological developments of X-ray computed tomography over half a century: User’s influence on protocol optimization. Eur. J. Radiol. 2020;131:109261. doi: 10.1016/j.ejrad.2020.109261. - DOI - PubMed
    1. van der Bie J., van Straten M., Booij R., Bos D., Dijkshoorn M.L., Hirsch A., Sharma S.P., Oei E.H.G., Budde R.P. Photon-counting CT: Review of initial clinical results. Eur. J. Radiol. 2023;163:110829. doi: 10.1016/j.ejrad.2023.110829. - DOI - PubMed
    1. Flohr T., Petersilka M., Henning A., Ulzheimer S., Ferda J., Schmidt B. Photon-counting CT review. Phys. Med. 2020;79:126–136. doi: 10.1016/j.ejmp.2020.10.030. - DOI - PubMed
    1. Danielsson M., Persson M., Sjölin M. Photon-counting X-ray detectors for CT. Phys. Med. Biol. 2021;66:03TR01. doi: 10.1088/1361-6560/abc5a5. - DOI - PubMed