In Vivo Meso-Scale Whole-Brain Quantitative Imaging With Tailored MRF on the NexGen 7T Scanner
- PMID: 41472402
- DOI: 10.1002/mrm.70234
In Vivo Meso-Scale Whole-Brain Quantitative Imaging With Tailored MRF on the NexGen 7T Scanner
Abstract
Purpose: To push the speed and resolution limit of in vivo quantitative imaging and enable estimation of quantitative tissue parameters of subtle brain structures that were previously difficult to assess.
Methods: This study implemented an efficient quantitative imaging approach, 3D-SPI MRF, on the NexGen 7T scanner equipped with a high-performance head-only gradient and 96-channel receiver array. To address challenges associated with performing rapid mesoscale MRF on this system, acquisition and reconstruction mitigation methods were developed and incorporated into the MRF framework, including: (i) flip-angle-aware dictionary fitting to account for both B1 + inhomogeneity and voxel-specific RF frequency response, (ii) gradient imperfection corrections via Skope measurements that incorporates a new per-TR trajectory rewinder compensation, (iii) incorporation of rapid B1 + and B0 mappings into the MRF sequence, and (iv) high-temporal motion navigation.
Results: Whole-brain T1 and T2 maps were obtained at 560-μm isotropic resolution within 4 min, where ablation studies demonstrated the necessity of the various mitigation methods implemented in removing bias and artifacts. For comparison, MRF data were acquired using current state-of-the-art method but limited to typical whole-body gradient specifications to demonstrate that the proposed developments resulted in ∼3× shorter scan time while producing more accurate parameter maps. Data were also acquired at ∼3.8× smaller voxel size, 360-μm isotropic, using the developed technique, to achieve mesoscale multi-parameter quantitative mapping in vivo.
Conclusion: Tailored 3D MRF acquisition and reconstruction were developed to enable fast and accurate T1 and T2 mapping across the whole-brain at mesoscale resolution on the NexGen 7T scanner.
Keywords: MRF; high field; mesoscale; quantitative mapping.
© 2025 International Society for Magnetic Resonance in Medicine.
References
-
- N. Weiskopf, L. J. Edwards, G. Helms, S. Mohammadi, and E. Kirilina, “Quantitative Magnetic Resonance Imaging of Brain Anatomy and In Vivo Histology,” Nature Reviews Physics 3, no. 8 (2021): 570–588, https://doi.org/10.1038/s42254‐021‐00326‐1.
-
- A. R. Wearn, V. Nurdal, E. Saunders‐Jennings, et al., “T2 Heterogeneity: A Novel Marker of Microstructural Integrity Associated With Cognitive Decline in People With Mild Cognitive Impairment,” Alzheimer's Research & Therapy 12, no. 1 (2020): 105, https://doi.org/10.1186/s13195‐020‐00672‐9.
-
- C. Liao, X. Cao, S. S. Iyer, et al., “High‐Resolution Myelin‐Water Fraction and Quantitative Relaxation Mapping Using 3D ViSTa‐MR Fingerprinting,” Magnetic Resonance in Medicine 91, no. 6 (2024): 2278–2293, https://doi.org/10.1002/mrm.29990.
-
- F. Lüsebrink, H. Mattern, R. Yakupov, et al., “Comprehensive Ultrahigh Resolution Whole Brain in Vivo MRI Dataset as a Human Phantom,” Scientific Data 8, no. 1 (2021): 138, https://doi.org/10.1038/s41597‐021‐00923‐w.
-
- D. Ma, V. Gulani, N. Seiberlich, et al., “Magnetic Resonance Fingerprinting,” Nature 495, no. 7440 (2013): 187–192, https://doi.org/10.1038/nature11971.
Grants and funding
LinkOut - more resources
Full Text Sources
Miscellaneous