Mass airflow cabinet for control of airborne infection of laboratory rodents

Mass airflow cabinet for control of airborne infection of laboratory rodents

G J McGarrity et al. Appl Microbiol. 1973 Aug.

Abstract

A mass airflow cabinet for handling and housing of laboratory rodents has been developed and tested. The unit consists of a high-efficiency particulate air filter and uniform distribution of air at a vertical velocity of 19 cm per s. Animals are maintained without bedding in mesh-bottomed cages that rest on rollers for rotation inside the cabinet. There is an air barrier of 90 cm per s separating the cabinet air from room air. Sampling for airborne bacteria yielded an average of 0.03 colony-forming units (CFU) per ft(3) of air inside the cabinet, whereas 28.8 CFU per ft(3) was simultaneously detected outside the cabinet during housekeeping, a reduction of almost three logs. The efficiency of the air barrier was tested by aerosolization of T3 phage. When phage was aerosolized 5 cm outside the cabinet, no phage could be detected 5 cm inside when the fans were operating; with the fans off an average of 1.6 x 10(4) plaque-forming units (PFU) per ft(3) was detected in six tests. Aerosolization of phage inside the cabinet yielded an average of 9 x 10 PFU per ft(3) outside; an average of 4.1 x 10(6) PFU per ft(3) were detected with the fans not in operation, a reduction of more than four logs. In-use studies on effectiveness showed that the cabinet significantly reduced the incidence of mice originally titer-free to Reo-3 virus. Hemagglutination inhibition antibodies to Reo-3 were detected in 9/22 (42%) mice housed in a conventionally ventilated animal laboratory while no seroconversion was detected in any of 22 mice housed in the mass air flow cabinet in the same laboratory.

PubMed Disclaimer

References

    1. Bacteriol Rev. 1961 Mar;25:18-31 - PubMed
    1. J Am Vet Med Assoc. 1972 Dec 1;161(11):1557-67 - PubMed
    1. JAMA. 1968 Mar 18;203(12):1038-46 - PubMed
    1. Am J Public Health Nations Health. 1967 Oct;57(10):1824-36 - PubMed
    1. J Natl Cancer Inst. 1965 Mar;34:371-80 - PubMed

LinkOut - more resources