Anesthesia equipment malfunction: origins and clinical recognition
- PMID: 436069
- PMCID: PMC1819256
Anesthesia equipment malfunction: origins and clinical recognition
Abstract
Equipment malfunction is a problem of particular importance during anesthesia and resuscitation. A review of published reports shows that the most common clinical events involve endotracheal tubes, the inspired oxygen concentration, the volume of inspired anesthetic vapours and gases, and pressures in the breathing or ventilation system. It is concluded that protection of a patient from equipment malfunction depends on: (a) appropriate application of standards set by a national standards association; (b) careful evaluation of equipment prior to purchase; (c) comprehension of equipment function by the user; (d) conscientious routine servicing of all systems concerned with anesthesia and resuscitation, and checking after service and before clinical use; (e) preanesthesia testing of equipment, including the use of an oxygen analyser in the breathing circuit; (f) early inclusion of equipment malfunction in the differential diagnosis of events during anesthesia; and (g) rapid action that cannot present a new hazard to the patient to correct the results of apparatus malfunction.
Similar articles
-
The Thai Anesthesia Incident Monitoring Study (Thai AIMS) of anesthetic equipment failure/malfunction: an analysis of 1996 incident reports.J Med Assoc Thai. 2009 Nov;92(11):1442-9. J Med Assoc Thai. 2009. PMID: 19938735
-
The Thai Anesthesia Incidents Study (THAI Study) of anesthetic equipment failure/malfunction: a qualitative analysis for risk factors.J Med Assoc Thai. 2005 Nov;88 Suppl 7:S134-40. J Med Assoc Thai. 2005. PMID: 16858992
-
Malfunction of anesthesia equipment.Can Med Assoc J. 1979 Sep 8;121(5):521. Can Med Assoc J. 1979. PMID: 497944 Free PMC article. No abstract available.
-
[Future perspectives for equipment maintenance and servicing].Anaesthesist. 1987 Mar;36(3):126-31. Anaesthesist. 1987. PMID: 3296850 Review. German.
-
The use of filters with small infants.Respir Care Clin N Am. 2006 Jun;12(2):307-20. doi: 10.1016/j.rcc.2006.03.006. Respir Care Clin N Am. 2006. PMID: 16828697 Review.
Cited by
-
Flying blind in anesthesia: A safety concern.Saudi J Anaesth. 2014 Jul;8(3):434-6. doi: 10.4103/1658-354X.136650. Saudi J Anaesth. 2014. PMID: 25191208 Free PMC article.
-
An unexpected error in oxygen humidifier.Saudi J Anaesth. 2014 Oct;8(4):567-8. doi: 10.4103/1658-354X.140911. Saudi J Anaesth. 2014. PMID: 25422623 Free PMC article. No abstract available.
-
Profound hypercapnia due to disconnection within an anaesthetic machine.Can J Anaesth. 1987 Nov;34(6):622-6. doi: 10.1007/BF03010524. Can J Anaesth. 1987. PMID: 3677289
-
Hazard notice; improper connections and damaged outer tube in oxygen humidifier.Anesth Essays Res. 2014 Sep-Dec;8(3):416-7. doi: 10.4103/0259-1162.143177. Anesth Essays Res. 2014. PMID: 25886349 Free PMC article. No abstract available.
-
Anaesthesia breathing circuits.Can Anaesth Soc J. 1986 Jan;33(1):98-105. doi: 10.1007/BF03010918. Can Anaesth Soc J. 1986. PMID: 3081242 Review. No abstract available.
References
MeSH terms
LinkOut - more resources
Full Text Sources