Modified laminar flow biological safety cabinet
- PMID: 4420479
- PMCID: PMC365796
- DOI: 10.1128/am.28.4.647-650.1974
Modified laminar flow biological safety cabinet
Abstract
Tests are reported on a modified laminar flow biological safety cabinet in which the return air plenum that conducts air from the work area to the high efficiency particulate air filters is under negative pressure. Freon gas released inside the cabinet could not be detected outside by a freon gas detection method capable of detecting 10(-6) cc/s. When T3 bacteriophage was aerosolized 5 cm outside the front opening in 11 tests, no phage could be detected inside the cabinet with the motor-filter unit in operation. An average of 2.8 x 10(5) plaque-forming units (PFU)/ft(3) (ca. 0.028 m(3)) were detected with the motor-filter unit not in operation, a penetration of 0.0%. Aerosolization 5 cm inside the cabinet yielded an average of 10 PFU/ft(3) outside the cabinet with the motor-filter unit in operation and an average of 4.1 x 10(5) PFU/ft(3) with the motor-filter unit not in operation, a penetration of 0.002%. These values are the same order of effectiveness as the positive-pressure laminar flow biological safety cabinets previously tested. The advantages of the negative-pressure return plenum design include: (i) assurance that if cracks or leaks develop in the plenum it will not lead to discharge of contaminated air into the laboratory; and (ii) the price is lower due to reduced manufacturing costs.
Similar articles
-
Bacteriological testing of a modified laminar flow microbiological safety cabinet.Antonie Van Leeuwenhoek. 1982;48(4):373-81. doi: 10.1007/BF00418290. Antonie Van Leeuwenhoek. 1982. PMID: 7149701
-
Influence of crossdrafts on the performance of a biological safety cabinet.Appl Environ Microbiol. 1978 Aug;36(2):278-83. doi: 10.1128/aem.36.2.278-283.1978. Appl Environ Microbiol. 1978. PMID: 100057 Free PMC article.
-
Mass airflow cabinet for control of airborne infection of laboratory rodents.Appl Microbiol. 1973 Aug;26(2):167-72. doi: 10.1128/am.26.2.167-172.1973. Appl Microbiol. 1973. PMID: 4355261 Free PMC article.
-
Microbiological studies on the performance of a laminar airflow biological cabinet.Appl Microbiol. 1968 Jul;16(7):1086-92. doi: 10.1128/am.16.7.1086-1092.1968. Appl Microbiol. 1968. PMID: 4874462 Free PMC article.
-
Biological safety cabinetry.Clin Microbiol Rev. 1991 Apr;4(2):207-41. doi: 10.1128/CMR.4.2.207. Clin Microbiol Rev. 1991. PMID: 2070345 Free PMC article. Review.
Cited by
-
Class II (laminar flow) biological safety cabinet.J Clin Pathol. 1979 May;32(5):505-13. doi: 10.1136/jcp.32.5.505. J Clin Pathol. 1979. PMID: 469008 Free PMC article.
-
Bacteriological testing of a modified laminar flow microbiological safety cabinet.Antonie Van Leeuwenhoek. 1982;48(4):373-81. doi: 10.1007/BF00418290. Antonie Van Leeuwenhoek. 1982. PMID: 7149701
-
Validation of cross-contamination control in biological safety cabinet for biotech/pharmaceutical manufacturing process.Environ Sci Pollut Res Int. 2015 Dec;22(23):19264-72. doi: 10.1007/s11356-015-5091-5. Epub 2015 Aug 11. Environ Sci Pollut Res Int. 2015. PMID: 26257118
-
Influence of crossdrafts on the performance of a biological safety cabinet.Appl Environ Microbiol. 1978 Aug;36(2):278-83. doi: 10.1128/aem.36.2.278-283.1978. Appl Environ Microbiol. 1978. PMID: 100057 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources