Indirect measurement of instantaneous arterial blood pressure in the rat
- PMID: 495770
- DOI: 10.1152/ajpheart.1979.237.5.H632
Indirect measurement of instantaneous arterial blood pressure in the rat
Abstract
We devised a hydraulic servo-control system for indirect blood pressure measurement in the rat's tail, by which beat-to-beat systolic and diastolic blood pressure can be obtained. In this method the principle of "unloading vascular wall" proposed by Shirer (1962) is employed. The proposed system is composed of a transmittance photoelectric plethysmograph with an occluding cuff, a small diaphragm actuator for compressing and decompressing the segment by the hydraulic pressure, and an electromagnetic shaker driven by a volume servo circuit in accordance with the signal from the photoelectric plethysmograph. The plethysmographic signal is clamped at a proper value corresponding to the unload vascular volume by the instantaneous hydraulic servo control. The cuff pressure thus automatically controlled follows the intra-arterial pressure in the tail segment. The accuracy of this method was evaluated in comparison with direct measurement of blood pressure recorded simultaneously from 16 unanesthetized spontaneously hypertensive and normotensive rats. Close agreement between the simultaneous data from these two methods were observed.
Similar articles
-
[Vascular-volume compensation method for noninvasive measurement of instantaneous arterial blood pressure: its usefulness in psychological research].Shinrigaku Kenkyu. 1983 Aug;54(3):204-7. Shinrigaku Kenkyu. 1983. PMID: 6672377 Japanese.
-
Comparison of simultaneous measurements of blood pressure by tail-cuff and carotid arterial methods in conscious spontaneously hypertensive and Wistar-Kyoto rats.Clin Exp Hypertens. 2006 Jan;28(1):57-72. doi: 10.1080/10641960500386817. Clin Exp Hypertens. 2006. PMID: 16443565
-
Validity of an indirect tail-cuff method for determining systolic arterial pressure in unanesthetized normotensive and spontaneously hypertensive rats.J Lab Clin Med. 1971 Dec;78(6):957-62. J Lab Clin Med. 1971. PMID: 5131859 No abstract available.
-
[Noninvasive, continuous measurement of finger artery pressure with the servo-plethysmo-manometer Finapres].Herz. 1989 Aug;14(4):251-9. Herz. 1989. PMID: 2676814 Review. German.
-
Criteria for set point estimation in the volume clamp method of blood pressure measurement.Physiol Res. 1992;41(1):5-10. Physiol Res. 1992. PMID: 1610779 Review.
Cited by
-
Noninvasive method for estimating the mean capillary pressure and pre- and postcapillary resistance ratio in human fingers.Med Biol Eng Comput. 1986 Nov;24(6):585-90. doi: 10.1007/BF02446260. Med Biol Eng Comput. 1986. PMID: 3657314 No abstract available.
-
Evaluation of the arterial blood pressure of dogs by two noninvasive methods.Can J Comp Med. 1985 Oct;49(4):419-23. Can J Comp Med. 1985. PMID: 4075242 Free PMC article.
-
Ambulatory monitoring of indirect beat-to-beat arterial pressure in human fingers by a volume-compensation method.Med Biol Eng Comput. 1991 Jan;29(1):55-62. doi: 10.1007/BF02446296. Med Biol Eng Comput. 1991. PMID: 2016921
-
Noninvasive automatic monitoring of instantaneous arterial blood pressure using the vascular unloading technique.Med Biol Eng Comput. 1983 Sep;21(5):557-65. doi: 10.1007/BF02442380. Med Biol Eng Comput. 1983. PMID: 6633005 No abstract available.
-
Noninvasive measurement of arterial elasticity in various human limbs.Med Biol Eng Comput. 1988 Nov;26(6):641-6. doi: 10.1007/BF02447504. Med Biol Eng Comput. 1988. PMID: 3256758 No abstract available.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources