Membrane vs bubble oxygenators: a prospective study of 52 patients
- PMID: 496630
- DOI: 10.1001/archsurg.1979.01370350087009
Membrane vs bubble oxygenators: a prospective study of 52 patients
Abstract
Controversy continues about the oxygenator preferable for cardiopulmonary bypass (CPB). This prospective study was undertaken in 52 patients undergoing coronary bypass surgery. Oxygenators were alternated each case between model Q-100, Bentley Laboratories Inc, Irvine, Calif, and Travenol Membrane Oxygenator (TMO), Travenol Laboratories Inc, Deerfield, Ill. The Q-100 group required higher CPB O2 flows, but PO2 levels during CPB were similar for both groups. Heparin sodium dosage and activated clotting, bleeding, prothrombin, and partial thromboplastin times were identical in both groups. Blood loss and platelet reduction after CPB were also similar. Postoperative complications in the Q-100 group included one myocardial infarction, and one neurological problem. The TMO group had no myocardial infarction and one neurological problem. The membrane oxygenator took nine minutes longer to set up and was $63 more expensive to purchase. Blood trauma during CPB was less with the membrane oxygenator (lower plasma hemoglobin level), but we conclude that both oxygenators performing adequately during clinical use in open heart surgery.
Similar articles
-
Membrane versus bubble oxygenator for cardiac operations. A prospective randomized study.J Thorac Cardiovasc Surg. 1980 Jul;80(1):111-22. J Thorac Cardiovasc Surg. 1980. PMID: 6966720 Clinical Trial.
-
Comparison of bubble and membrane oxygenators in short and long perfusions.J Thorac Cardiovasc Surg. 1979 Nov;78(5):655-66. J Thorac Cardiovasc Surg. 1979. PMID: 491720
-
Pressure drop, shear stress, and activation of leukocytes during cardiopulmonary bypass: a comparison between hollow fiber and flat sheet membrane oxygenators.Artif Organs. 2000 Jan;24(1):43-8. doi: 10.1046/j.1525-1594.2000.06351.x. Artif Organs. 2000. PMID: 10677156
-
Membrane oxygenators: current developments in design and application.J Biomed Eng. 1988 Nov;10(6):541-7. doi: 10.1016/0141-5425(88)90113-6. J Biomed Eng. 1988. PMID: 3070171 Review.
-
Development of the oxygenator: past, present, and future.J Artif Organs. 2004;7(3):111-20. doi: 10.1007/s10047-004-0268-6. J Artif Organs. 2004. PMID: 15558331 Review.
Cited by
-
Antiplatelet drugs: mechanisms and risks of bleeding following cardiac operations.Int J Angiol. 2011 Mar;20(1):1-18. doi: 10.1055/s-0031-1272544. Int J Angiol. 2011. PMID: 22532765 Free PMC article.
-
In vivo survival of red blood cells processed by a bubble or membrane oxygenator during cardiopulmonary bypass surgery.Tex Heart Inst J. 1986 Jun;13(2):247-51. Tex Heart Inst J. 1986. PMID: 15227369 Free PMC article.
-
Extracorporeal Hyperoxygenation Therapy (EHT) for Carbon Monoxide Poisoning: In-Vitro Proof of Principle.Membranes (Basel). 2021 Dec 31;12(1):56. doi: 10.3390/membranes12010056. Membranes (Basel). 2021. PMID: 35054581 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources