Spectral analysis of the normal electrocardiogram in children and adults
Abstract
We performed spectral analysis of the X-lead Frank VCG from six normal adults (Group A) and eight normal children (Group B) using a Fast Fourier Transform. The Group A magnitudes were consistently and significantly lower than Group B in the 20--90 Hz region. Similarly, the -40 dB point was 75 Hz for Group A and 107 Hz for Group B. The P waves of the two groups had similar spectra, as did their T waves. When the entire cardiac cycle was analyzed, there was again a consistent and significant difference in the frequencies from 25--80 Hz. Despite the significant differences in the Group A and B spectra, the vast majority of the information contained in the ECG signal was confined to frequencies below 100 Hz. Therefore, a bandwidth of 100 Hz is probably adequate for accurate recording of the ECG signal from children, as well as from adults.
Similar articles
-
Vectorcardiograms synthesized from 12-lead ECGs: a new method applied in 1792 healthy children.Pediatr Cardiol. 1994 Jan-Feb;15(1):21-6. doi: 10.1007/BF00797001. Pediatr Cardiol. 1994. PMID: 8115267
-
Diagnostic implications of spectral and temporal analysis of the entire cardiac cycle in patients with ventricular tachycardia.Circulation. 1991 May;83(5):1637-48. doi: 10.1161/01.cir.83.5.1637. Circulation. 1991. PMID: 2022021
-
[Development of computer assisted VCG analysis system and its clinical application (author's transl)].Kokyu To Junkan. 1981 Feb;29(2):153-7. Kokyu To Junkan. 1981. PMID: 7255968 Japanese. No abstract available.
-
Vectorcardiographic diagnostic & prognostic information derived from the 12-lead electrocardiogram: Historical review and clinical perspective.J Electrocardiol. 2015 Jul-Aug;48(4):463-75. doi: 10.1016/j.jelectrocard.2015.05.002. Epub 2015 May 8. J Electrocardiol. 2015. PMID: 26027545 Review.
-
A review of beat-to-beat vectorcardiographic (VCG) parameters for analyzing repolarization variability in ECG signals.Biomed Tech (Berl). 2016 Feb;61(1):3-17. doi: 10.1515/bmt-2015-0005. Biomed Tech (Berl). 2016. PMID: 25992510 Review.
Cited by
-
Frank vectorcardiograms during the first year of life: a longitudinal study of 62 healthy children.Pediatr Cardiol. 1986;7(3):141-6. doi: 10.1007/BF02424987. Pediatr Cardiol. 1986. PMID: 3808992
-
Spectral analysis of the normal electrocardiogram of rabbits.Basic Res Cardiol. 1985 Jul-Aug;80(4):353-6. doi: 10.1007/BF01908178. Basic Res Cardiol. 1985. PMID: 4051938
-
Suitable analyzed signal duration and criteria for abnormal signal-averaged electrocardiogram on frequency-domain analysis in patients with Duchenne's muscular dystrophy.Pediatr Cardiol. 1996 Mar-Apr;17(2):91-6. doi: 10.1007/BF02505090. Pediatr Cardiol. 1996. PMID: 8833493 Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Miscellaneous