Continuous comparison of in vitro and in vivo calibrated transcutaneous oxygen tension with arterial oxygen tension in infants
- PMID: 534709
Continuous comparison of in vitro and in vivo calibrated transcutaneous oxygen tension with arterial oxygen tension in infants
Abstract
Transcutaneous PO2 simultaneously recorded by Drager and Radiometer electrodes on the abdominal skin was compared for six-hour periods with aortic PaO2 recorded by a Searle intravascular oxygen electrode. Ten newborn infants with serious respiratory illnesses, six of whom needed mechanical ventilation, were studied. The skin electrodes were heated to 44 degrees C and calibrated first in vitro, and then in vivo against the infant's PaO2. The results showed that 1) after in vivo calibration both skin electrodes gave an accurate estimate of PaO2 for six hours without resiting of the electrodes; 2) the Radiometer electrode gave as satisfactory an estimate of PaO2 after in vitro as after in vivo calibration; 3) the Drager electrode gave a significantly less accurate estimate of PaO2 after in vitro than after in vivo calibration; 4) no evidence suggesting that peripheral vasoconstriction interfered with the accuracy of estimation of PaO2 by the skin electrodes was found.
Similar articles
-
Comparison of transcutaneous oxygen tension with arterial oxygen tension in newborn infants with severe respiratory illnesses.Pediatrics. 1978 Nov;62(5):692-7. Pediatrics. 1978. PMID: 724311
-
Continuous monitoring of arterial oxygen tension using a catheter-tip polarographic electrode in infants.Pediatrics. 1976 Feb;57(2):244-50. Pediatrics. 1976. PMID: 1250658
-
Continuous comparison of transcutaneous and arterial oxygen tension in newborn infants with respiratory illnesses.Acta Anaesthesiol Scand Suppl. 1978;68:91-7. doi: 10.1111/j.1399-6576.1978.tb01400.x. Acta Anaesthesiol Scand Suppl. 1978. PMID: 279208
-
Transcutaneous oxygen tension measurements in newborn infants.Pediatrics. 1975 Feb;55(2):232-5. Pediatrics. 1975. PMID: 1090895 Review.
-
Blood flow rate, temperature, oxygen tension and consumption in the skin of adults measured by a heated microcathode oxygen electrode.Dan Med Bull. 1988 Aug;35(4):322-34. Dan Med Bull. 1988. PMID: 3048920 Review.