Is the Bain breathing circuit the future anesthesia system? An evaluation
- PMID: 556914
- DOI: 10.1213/00000539-197701000-00020
Is the Bain breathing circuit the future anesthesia system? An evaluation
Abstract
The Bain breathing circuit, a modified Mapleson D system, was evaluated with regard to oxygenation and CO2 elimination under controlled conditions and compared with the presently popular semiclosed breathing circuit (SCBC) with CO2 absorber. The authors demonstrated that the Bain system compares favorably with the SCBC in regard to oxygenation of manually ventilated patients with fresh gas inflow of 70 ml/kg/min while maintaining a PaCO2 at a mean of 38 torr versus an SCBC mean of 32 torr. The authors were impressed with the clinical simplicity, efficiency, and versatility of the Bain system and believe that it will play a major role in the future of anesthesia-machine design.
Similar articles
-
The Bain anaesthetic system. An assessment during controlled ventilation.Anaesthesia. 1976 Mar;31(2):247-56. doi: 10.1111/j.1365-2044.1976.tb11798.x. Anaesthesia. 1976. PMID: 937671
-
Reduction of fresh gas flow requirements by a circle-modified bain breathing circuit.J Clin Monit. 1991 Jan;7(1):49-55. doi: 10.1007/BF01617899. J Clin Monit. 1991. PMID: 1900324
-
The Bain anaesthetic circuit.S Afr Med J. 1978 Jun 3;53(22):904-6. S Afr Med J. 1978. PMID: 356289
-
[Blood gas levels using spontaneous breathing and the Bain circuit. Our experience during head surgery (author's transl)].Anesth Analg (Paris). 1981;38(9-10):505-11. Anesth Analg (Paris). 1981. PMID: 6800285 French.
-
Fresh gas flow in coaxial Mapleson A and D circuits during spontaneous breathing.Acta Anaesthesiol Scand. 1986 Oct;30(7):588-93. doi: 10.1111/j.1399-6576.1986.tb02481.x. Acta Anaesthesiol Scand. 1986. PMID: 3101384
Cited by
-
The "D circle": closed-circuit operation of the Bain circuit.Can Anaesth Soc J. 1979 Sep;26(5):420-3. doi: 10.1007/BF03006458. Can Anaesth Soc J. 1979. PMID: 290406
-
How rebreathing anaesthetic systems control PaCO2: studies with a mechanical and a mathematical model.Can Anaesth Soc J. 1978 Mar;25(2):117-21. doi: 10.1007/BF03005067. Can Anaesth Soc J. 1978. PMID: 638822
-
Carbon dioxide output and elimination in children under anaesthesia.Can Anaesth Soc J. 1977 Sep;24(5):533-9. doi: 10.1007/BF03005526. Can Anaesth Soc J. 1977. PMID: 907892
-
PEEP and the Bain circuit.Can Anaesth Soc J. 1981 Sep;28(5):467-70. doi: 10.1007/BF03010358. Can Anaesth Soc J. 1981. PMID: 7025997
-
Spontaneous respiration with the Bain breathing circuit.Can Anaesth Soc J. 1978 Jan;25(1):30-5. doi: 10.1007/BF03006780. Can Anaesth Soc J. 1978. PMID: 272223 No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources