Experimental study with the Scurasil ureteral prosthesis
- PMID: 591271
Experimental study with the Scurasil ureteral prosthesis
Abstract
The latest generation of Scurasil (dimethylpolisoxane) ureteral prosthesis was used in six dogs as a model for internal urinary diversion in patients with metastatic disease. It was found to cause minimal reaction but carried a high complication rate, including stone formation, end-stage pyelonephritis, and hydronephrosis. Although technically easy to use, modifications in design are necessary prior to use in humans.
Similar articles
-
Ureteral stenting.Semin Urol. 1984 Aug;2(3):180-6. Semin Urol. 1984. PMID: 6385191 No abstract available.
-
[Silicone ureteral prostheses: results of an experimental study of tolerance and surface conditions over six years, and clinical use for four years (author's transl)].J Chir (Paris). 1981 Jan;118(1):1-13. J Chir (Paris). 1981. PMID: 7204509 French.
-
Permanent silastic ureteral prostheses.Surg Forum. 1962;13:501-3. Surg Forum. 1962. PMID: 13971858 No abstract available.
-
Ureteral replacements.Surg Gynecol Obstet. 1972 Nov;135(5):801-10. Surg Gynecol Obstet. 1972. PMID: 4563512 Review. No abstract available.
-
Ureteral prostheses.Semin Urol. 1984 Aug;2(3):158-66. Semin Urol. 1984. PMID: 6237406 Review. No abstract available.