Errors in microdensitometry
- PMID: 6166593
- DOI: 10.1007/BF01006883
Errors in microdensitometry
Abstract
Microdensitometric errors can originate in the instrument, in the specimen or in the human operator. Instrumental sources of systematic error mostly reduce the apparent integrated absorbance, especially of relatively small and highly absorbing objects. They can be assessed, minimized or eliminated by available techniques, but with modern apparatus are in general important only if results of high accuracy are required. Instrument errors include: (a) distributional error, due to the use of too large a measuring spot or the specimen being out of focus; (b) glare (stray light), due mainly to multiple reflections in the microscope objective; (c) monochromator error (the use of insufficiently pure light); (d) calibration errors; and (e) errors resulting from lack of photometric linearity, or the specimen absorbance exceeding the measuring range of the instrument. Specimen errors, including the problems of specificity and stoichiometry, are now the most important obstacles to a wider use of microdensitometry. The following selected topics are briefly discussed: fading; rate of staining; Beer's law deviations and the microdensitometry of opaque particles. Human errors include faulty logic, and failing to attempt an investigation because of anticipated difficulties which are in fact exaggerated or imaginary. The significance of microdensitometric results should, in general, be assessed by biological criteria rather than merely statistically; the use is urged of appropriate internal biological controls and standards wherever possible.
Similar articles
-
Scanning microdensitometry of objects small relative to the wavelength of light.J Histochem Cytochem. 1982 Oct;30(10):1040-50. doi: 10.1177/30.10.7130668. J Histochem Cytochem. 1982. PMID: 7130668
-
Aspects of scanning microdensitometry. III. The monochromator system.J Microsc. 1975 Sep;105(1):33-56. doi: 10.1111/j.1365-2818.1975.tb04035.x. J Microsc. 1975. PMID: 1243150
-
Apparent anomalies in nuclear feulgen-DNA contents. Role of systematic microdensitometric errors.J Cell Biol. 1976 Oct;71(1):68-88. doi: 10.1083/jcb.71.1.68. J Cell Biol. 1976. PMID: 61968 Free PMC article.
-
A user's guide for avoiding errors in absorbance image cytometry: a review with original experimental observations.Histochem J. 1994 Jan;26(1):1-19. Histochem J. 1994. PMID: 7513318 Review.
-
Measuring and reporting errors in surgical pathology. Lessons from gynecologic cytology.Am J Clin Pathol. 2001 Mar;115(3):338-41. doi: 10.1309/M2XP-3YJA-V6E2-QD9P. Am J Clin Pathol. 2001. PMID: 11242788 Review.
Cited by
-
The out-of-range error in microdensitometry.Histochem J. 1986 Jul;18(7):397-8. doi: 10.1007/BF01675221. Histochem J. 1986. PMID: 3759498 No abstract available.
-
Reaction rate measurements of proteases and glycosidases with chromogenic methods.Histochem J. 1989 Sep-Oct;21(9-10):535-44. doi: 10.1007/BF01753354. Histochem J. 1989. PMID: 2687213
-
Quantitative succinate dehydrogenase histochemistry in the hippocampus of aged rats.Histochemistry. 1988;88(3-6):299-307. doi: 10.1007/BF00570287. Histochemistry. 1988. PMID: 3366636
-
Cytochemical properties of euchromatin and heterochromatin.Histochem J. 1983 Aug;15(8):739-51. doi: 10.1007/BF01003338. Histochem J. 1983. PMID: 6194139 Review.
-
Quantitative enzyme cytochemistry during human macrophage development.J Anat. 1993 Aug;183 ( Pt 1)(Pt 1):97-101. J Anat. 1993. PMID: 8270480 Free PMC article.