A comparison of serum electrolyte results by the Flame III Auto-Analyser and a Flame IV Auto-Analyser which utilises curve regeneration with a data acquisition system
- PMID: 630699
- DOI: 10.1016/s0009-9120(78)80039-3
A comparison of serum electrolyte results by the Flame III Auto-Analyser and a Flame IV Auto-Analyser which utilises curve regeneration with a data acquisition system
Abstract
Results obtained with a Technicon Flame Photometer Model IV Auto-Analyzer, which utilizes curve regeneration in conjunction with a mini-computer for the continuous-flow determination of serum electrolytes, is presented. The system, featuring optimal flow conditions, subjects sample peaks to a mathematical treatment to effectively remove exponential deformation. The evaluation, undertaken after 2 months of operation in the routine clinical laboratory, utilizes the Flame Photometer Model III as the "referee" procedure. The comparison includes a statistical analysis of quality control reference material, simultaneously determined patient data and an evaluation of repeated patient data. The system gave comparable results to those obtained for the Flame Photometer Model III and provides the additional benefits of increased sample throughput, decreased reagent comsumption and sample interaction and gives reductions in the clerical tasks associated with the retrieval of data from Auto-Analyzer systems.
Similar articles
-
Application of curve regeneration to the determination of electrolytes by the Flame IV Auto-Analyser using a mini-computer.Clin Biochem. 1978 Feb;11(1):16-22. doi: 10.1016/s0009-9120(78)80038-1. Clin Biochem. 1978. PMID: 630698
-
A comparison of serum electrolyte results by Technicon's Flame III and Flame IV AutoAnalyser system.Clin Biochem. 1978 Feb;11(1):10-5. doi: 10.1016/s0009-9120(78)80037-x. Clin Biochem. 1978. PMID: 630697
-
An inexpensive system of automated flame photometry of serum sodium and potassium.Ann Clin Biochem. 1974 Nov;11(6):238-41. doi: 10.1177/000456327401100169. Ann Clin Biochem. 1974. PMID: 4460848
-
Comparison of serum sodium and chloride results for Flame IV-Auto Analyzer II, SMAC, Ektachem 400, and Nova 4 clinical analyzer systems.Clin Biochem. 1988 Jun;21(3):167-72. doi: 10.1016/0009-9120(88)90005-7. Clin Biochem. 1988. PMID: 3390906
-
[Determination of electrolytes in serum and plasma].Wien Klin Wochenschr Suppl. 1992;192:37-41. Wien Klin Wochenschr Suppl. 1992. PMID: 1502823 Review. German.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources