Use of leukocyte esterase-nitrate activity as predictive assays of significant bacteriuria
- PMID: 6358253
- PMCID: PMC272876
- DOI: 10.1128/jcm.18.5.1256-1257.1983
Use of leukocyte esterase-nitrate activity as predictive assays of significant bacteriuria
Abstract
The present study evaluated the usefulness of the 1-min leukocyte esterase-nitrite tests in a tertiary-care hospital as a screening procedure to detect significant bacteriuria and correlated the findings with culture results. A total of 531 urine samples were reviewed, of which 484 were evaluated. Of the evaluated samples, 113 positive cultures (23.4%) were found, of which 93 (82.3%) were detected by leukocyte esterase-nitrite tests. In addition, 365 of 371 (98.4%) urine samples with negative bacterial cultures were negative in leukocyte esterase and nitrite tests.