Immunofluorescence measurement in a flow cytometer using low-power helium-neon laser excitation

Free article

Immunofluorescence measurement in a flow cytometer using low-power helium-neon laser excitation

H M Shapiro et al. Cytometry. 1983 Nov.
Free article

Abstract

Helium-neon lasers are economical and efficient light sources; their utility in flow cytometry to date has been limited by the lack of fluorescent probes that can be excited at 633 nm. Allophycocyanin (APC), a highly fluorescent phycobiliprotein, can be used as an antibody label and has spectral characteristics suitable for use with He-Ne lasers; we undertook to resolve whether a low-power (7 mW) He-Ne laser could provide sufficient excitation to permit flow cytometric detection of APC-labeled antibodies on cell surfaces. We made an APC conjugate of monoclonal antibody 4F2, which reacts with an antigen abundant on the surfaces of activated human T-lymphocytes; APC-4F2 was used to stain blood mononuclear cells that had been cultured with and without phytohemagglutinin (PHA). Cells so stained were examined in a flow cytometer with orthogonal illumination at 633 nm from a 7 mW He-Ne laser; antibody-bearing cells were detectable by fluorescence emission above 665 nm. Cells from the same cultures were stained with fluorescein-labeled 4F2 antibody and examined in a flow cytometer with argon ion laser excitation at 488 nm. Percentages of antibody-bearing cells determined from APC fluorescence and from fluorescein fluorescence were in good agreement. It thus appears that He-Ne lasers and APC-antibodies are usable for immunofluorescence measurements; the sensitivity attainable with this technique remains to be determined.

PubMed Disclaimer

Similar articles

Cited by

Publication types

LinkOut - more resources