The measurement of total and unsaturated iron-binding capacity in serum
- PMID: 638076
- DOI: 10.1111/j.1365-2141.1978.tb01044.x
The measurement of total and unsaturated iron-binding capacity in serum
Abstract
An inter-laboratory investigation into the assay of serum iron binding capacity by the magnesium carbonate adsorption method was undertaken. The source of MgCO3, storage temperature of whole serum, lyophilization and storage temperature of lyophilized serum were identified as sources of inter-laboratory variation. Protocols for the clinical assay of UIBC and TIBC are given which have achieved satisfactory inter-laboratory precision.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources