A human calorimeter for the direct and indirect measurement of 24 h energy expenditure
- PMID: 638125
- DOI: 10.1079/bjn19780071
A human calorimeter for the direct and indirect measurement of 24 h energy expenditure
Abstract
1. A calorimeter for the continuous measurement of heat production and heat loss in the human subject, for at least 24 h, is described. The calorimeter operated on the heat-sink principle for direct calorimetry and an open-circuit system for indirect calorimetry. 2. Sensible heat loss was measured using a water-cooled heat exchanger, and the temperature of water entering the heat exchanger was controlled to maintain a mean temperature gradient of zero across the chamber walls. 3. Evaporative heat loss was determined from ingoing and outgoing wet-and-dry bulb temperatures and air flow-rates. 4. Problems associated with the calculation of evapoative heat loss and the estimation of the volume of incoming air in open-circuit systems are considered. 5. The calibration, limits of accuracy, sources of error and experiments with subjects are discussed.