Air versus skin temperature servocontrol of infant incubators
Abstract
Air temperature servocontrol was compared with skin temperature servocontrol and manual control as methods for regulating the heat output of a single-walled incubator (Air-Shields C-86) (1) when optimally used in the laboratory and (2) when operated by staff nurses in the nursery. The subjects were eight premature infants with birth weights between 1.07 and 1.54 kg. When the three methods were used to produce neutral air and skin temperatures during 2-hour measurement periods in the laboratory, there were no differences in mean air, skin, or rectal temperature, metabolic heat production, or body heat loss. There were also no differences among the three methods in mean air, skin, or rectal temperature when used by the nurses in the nursery for periods of 24 hours. When incubator wall temperature is stable, air temperature servocontrol can be used as effectively as skin temperature servocontrol to operate infant incubators.
Similar articles
-
A double-walled incubator alters the partition of body heat loss of premature infants.Pediatr Res. 1983 Feb;17(2):135-40. doi: 10.1203/00006450-198302000-00011. Pediatr Res. 1983. PMID: 6402753
-
Effects of low humidity on small premature infants in servocontrol incubators. I. Decrease in rectal temperature.Biol Neonate. 1975;26(5-6):337-47. doi: 10.1159/000240747. Biol Neonate. 1975. PMID: 1174595
-
Infant incubators and radiant warmers.Early Hum Dev. 1983 Oct;8(3-4):351-75. doi: 10.1016/0378-3782(83)90018-x. Early Hum Dev. 1983. PMID: 6641579
-
Neutral temperature range in incubators: performance of equipment in current use and new developments.Crit Rev Biomed Eng. 1997;25(4-5):287-370. doi: 10.1615/critrevbiomedeng.v25.i4-5.10. Crit Rev Biomed Eng. 1997. PMID: 9505136 Review.
-
Mathematical modelling of thermoregulation processes for premature infants in closed convectively heated incubators.Comput Biol Med. 2015 Feb;57:159-72. doi: 10.1016/j.compbiomed.2014.11.021. Epub 2014 Dec 9. Comput Biol Med. 2015. PMID: 25557201 Review.
Cited by
-
Dynamics and complexity of body temperature in preterm infants nursed in incubators.PLoS One. 2017 Apr 27;12(4):e0176670. doi: 10.1371/journal.pone.0176670. eCollection 2017. PLoS One. 2017. PMID: 28448569 Free PMC article.
-
Incubator temperature control: effects on the very low birthweight infant.Arch Dis Child. 1985 Oct;60(10):902-7. doi: 10.1136/adc.60.10.902. Arch Dis Child. 1985. PMID: 4062342 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources