In vivo response time of transcutaneous oxygen measurement to changes in inspired oxygen in normal adults
- PMID: 6692181
- DOI: 10.1007/BF03011489
In vivo response time of transcutaneous oxygen measurement to changes in inspired oxygen in normal adults
Abstract
We examined the in vivo response time of transcutaneous oxygen measurement (PtcO2) to a step increase (100 per cent oxygen) or decrease (air) of inspired oxygen in five normal adults (age 27 +/- SD 4.5 years). The PtcO2 electrode was placed on the upper chest (sensor temperature 44 degrees C). The mean +/- SEM 63 and 95 per cent response times of PtcO2 to a step increase in inspired oxygen were 2.9 +/- 0.2 and 6.8 +/- 0.2 and to a step decrease were 2.4 +/- 0.2 and 5.1 +/- 0.3 minutes. The 63 and 95 per cent response times for the pulmonary washin of oxygen were 1.0 and 2.6 minutes and washout were 0.8 and 2.3 minutes. We conclude that the in vivo response time of the PtcO2 is reproducible but substantially longer than that of the pulmonary washin and washout times and inadequate to monitor rapid changes in arterial oxygenation during anaesthesia, perioperative period or sleep in adults.
Similar articles
-
Transcutaneous oxygen measurement during one-lung anaesthesia.Acta Anaesthesiol Scand. 1984 Jun;28(3):241-4. doi: 10.1111/j.1399-6576.1984.tb02051.x. Acta Anaesthesiol Scand. 1984. PMID: 6430006
-
Noninvasive monitoring in emergency resuscitation.Ann Emerg Med. 1986 Dec;15(12):1434-6. doi: 10.1016/s0196-0644(86)80937-4. Ann Emerg Med. 1986. PMID: 3777616
-
Transcutaneous oxygen measurement in adult intensive care.Acta Anaesthesiol Scand. 1979 Dec;23(6):534-44. doi: 10.1111/j.1399-6576.1979.tb01484.x. Acta Anaesthesiol Scand. 1979. PMID: 545990
-
Noninvasive monitoring of oxygen and carbon dioxide.Clin Chest Med. 1989 Jun;10(2):239-54. Clin Chest Med. 1989. PMID: 2661121 Review.
-
Transcutaneous PO2 measurement.Can Anaesth Soc J. 1984 Nov;31(6):664-77. doi: 10.1007/BF03008765. Can Anaesth Soc J. 1984. PMID: 6388743 Review.
Cited by
-
Comparison of tissue oxygenation achieved breathing oxygen from a demand valve with four different mask configurations.Diving Hyperb Med. 2018 Dec 24;48(4):209-217. doi: 10.28920/dhm48.4.209-217. Diving Hyperb Med. 2018. PMID: 30517952 Free PMC article.
-
Effect of Incentive Spirometry on Postoperative Hypoxemia and Pulmonary Complications After Bariatric Surgery: A Randomized Clinical Trial.JAMA Surg. 2017 May 1;152(5):422-428. doi: 10.1001/jamasurg.2016.4981. JAMA Surg. 2017. PMID: 28097332 Free PMC article. Clinical Trial.
-
Noninvasive monitoring of oxygenation during one-lung ventilation: a comparison of transcutaneous oxygen tension measurement and pulse oximetry.J Clin Monit. 1987 Apr;3(2):90-5. doi: 10.1007/BF00858356. J Clin Monit. 1987. PMID: 3585438
-
Comparison of tissue oxygenation achieved breathing oxygen using different delivery devices and flow rates.Diving Hyperb Med. 2020 Mar 31;50(1):34-42. doi: 10.28920/dhm50.1.34-42. Diving Hyperb Med. 2020. PMID: 32187616 Free PMC article.
-
The influence of upper respiratory tract surgery on respiratory function evaluated by oxygen saturation.J Anesth. 1991 Oct;5(4):327-30. doi: 10.1007/s0054010050327. J Anesth. 1991. PMID: 15278599