A method for correcting bone induced artifacts in computed tomography scanners
- PMID: 670461
- DOI: 10.1097/00004728-197801000-00017
A method for correcting bone induced artifacts in computed tomography scanners
Abstract
A method that corrects for artifacts in X-ray transmission scanners created by the alteration of the energy spectrum by bone is described. The method involves two reconstructions of the pictures: the first establishes the approximate distribution of bone and the second enables the artifacts to be eliminated. The first reconstruction is needed to establish the total amount of bone along each ray. Assuming that the X-ray energy spectrum is known, it is shown how the line integrals can be corrected. The method was tested on several mathematical phantoms as well as on human head data. The nonlinear nature of the spectral artifacts is shown to produce streaks in certain phantoms.
Similar articles
-
A method for simultaneous correction of spectrum hardening artifacts in CT images containing both bone and iodine.Med Phys. 1997 Oct;24(10):1629-34. doi: 10.1118/1.597970. Med Phys. 1997. PMID: 9350717
-
A new software correction approach to volume averaging artifacts in CT.Comput Radiol. 1986 Mar-Jun;10(2-3):87-98. doi: 10.1016/0730-4862(86)90050-8. Comput Radiol. 1986. PMID: 3754803
-
A full-spectral Bayesian reconstruction approach based on the material decomposition model applied in dual-energy computed tomography.Med Phys. 2013 Nov;40(11):111916. doi: 10.1118/1.4820478. Med Phys. 2013. PMID: 24320449
-
A comparison of maxillofacial CBCT and medical CT.Atlas Oral Maxillofac Surg Clin North Am. 2012 Mar;20(1):1-17. doi: 10.1016/j.cxom.2011.12.008. Atlas Oral Maxillofac Surg Clin North Am. 2012. PMID: 22365427 Review. No abstract available.
-
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.Semin Nucl Med. 2003 Jul;33(3):166-79. doi: 10.1053/snuc.2003.127307. Semin Nucl Med. 2003. PMID: 12931319 Review.
Cited by
-
[An adaptive CT metal artifact reduction algorithm that combines projection interpolation and physical correction].Nan Fang Yi Ke Da Xue Xue Bao. 2022 Jun 20;42(6):832-839. doi: 10.12122/j.issn.1673-4254.2022.06.06. Nan Fang Yi Ke Da Xue Xue Bao. 2022. PMID: 35790433 Free PMC article. Chinese.
-
Cone-Beam CT of Traumatic Brain Injury Using Statistical Reconstruction with a Post-Artifact-Correction Noise Model.Proc SPIE Int Soc Opt Eng. 2015 Feb 21;9412:941207. doi: 10.1117/12.2082075. Proc SPIE Int Soc Opt Eng. 2015. PMID: 26300578 Free PMC article.
-
Image Quality Evaluation for Brain Soft Tissue in Neuroendovascular Treatment by Dose-Reduction Mode of Dual-Axis "Butterfly" Scan.AJNR Am J Neuroradiol. 2025 Feb 3;46(2):285-292. doi: 10.3174/ajnr.A8472. AJNR Am J Neuroradiol. 2025. PMID: 39179299
-
Dynamic fluence field modulation in computed tomography using multiple aperture devices.Phys Med Biol. 2019 May 21;64(10):105024. doi: 10.1088/1361-6560/ab155e. Phys Med Biol. 2019. PMID: 30939459 Free PMC article.
-
Dose reduction using a dynamic, piecewise-linear attenuator.Med Phys. 2014 Feb;41(2):021910. doi: 10.1118/1.4862079. Med Phys. 2014. PMID: 24506631 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources