Correction for beam hardening in computed tomography
- PMID: 672814
- DOI: 10.1118/1.594429
Correction for beam hardening in computed tomography
Abstract
Corrections for beam-hardening artifacts in computed tomography can be made by using a model which assumes that water and bone mineral are the only constituents of tissue. With this model, a correction factor for the measured transmission values can be calculated such that the reconstructed attenuation coefficients have values corresponding to a monoenergetic source of known energy. Systematic errors in the uncorrected attenuation coefficients, which may be 5%, can be reduced to less than 1% if corrected transmission values are used.