Oxygen consumption and insensible water loss in premature infants under radiant heaters
- PMID: 6773022
Oxygen consumption and insensible water loss in premature infants under radiant heaters
Abstract
Oxygen consumption ((Vo2), carbon dioxide production (Vco2), and insensible water loss (IWL) were measured simultaneously in nine nondistressed, appropriately grown, premature infants less than 2 weeks old, nursed in a conventional, blow-warmed incubator, and were compared with measurements made on the same infants under a radiant heater. The infants had a pronounced increase (148% on average) in IWL when under the radiant heater (P < .001) whereas Vo2 increased by only 4.6% (P = .073). Abdominal skin temperature (servocontrolled to maintain 36.5 C) and esophageal temperature were the same under both conditions, but ambient air temperature was 0.7 C higher in the incubator (P < 05). Although a positive correlation was found between the increase in IWL and the change in Vo2 (r = .75, P < .01), the large increase in IWL (and, therefore, evaporative heat loss) under the radiant heater is out of proportion to, and cannot be accounted for, by the change in metabolic heat production. The heat transfer processes involved in maintaining body temperature constant under these conditions require further study.
Similar articles
-
Oxygen consumption and temperature control of premature infants in a double-wall incubator.Pediatrics. 1981 Jul;68(1):93-8. Pediatrics. 1981. PMID: 7243514 Clinical Trial.
-
Resting oxygen consumption of premature infants covered with a plastic thermal blanket.Pediatrics. 1979 Apr;63(4):547-51. Pediatrics. 1979. PMID: 440864
-
A double-walled incubator alters the partition of body heat loss of premature infants.Pediatr Res. 1983 Feb;17(2):135-40. doi: 10.1203/00006450-198302000-00011. Pediatr Res. 1983. PMID: 6402753
-
Radiant energy and insensible water loss in the premature newborn infant nursed under a radiant warmer.Clin Perinatol. 1982 Oct;9(3):483-503. Clin Perinatol. 1982. PMID: 6761034 Review.
-
Insensible water loss in newborn infants.Clin Perinatol. 1982 Oct;9(3):453-82. Clin Perinatol. 1982. PMID: 6761033 Review.
Cited by
-
Radiant warmers versus incubators for regulating body temperature in newborn infants.Cochrane Database Syst Rev. 2000;2003(2):CD000435. doi: 10.1002/14651858.CD000435. Cochrane Database Syst Rev. 2000. Update in: Cochrane Database Syst Rev. 2002;(2):CD000435. doi: 10.1002/14651858.CD000435. PMID: 10796363 Free PMC article. Updated.
-
Assessing neonatal heat balance and physiological strain in newborn infants nursed under radiant warmers in intensive care with fentanyl sedation.Eur J Appl Physiol. 2014 Dec;114(12):2539-49. doi: 10.1007/s00421-014-2964-0. Epub 2014 Aug 13. Eur J Appl Physiol. 2014. PMID: 25115506
-
Open or closed incubators.Arch Dis Child. 1986 Feb;61(2):108-9. doi: 10.1136/adc.61.2.108. Arch Dis Child. 1986. PMID: 3954434 Free PMC article. No abstract available.
MeSH terms
Substances
LinkOut - more resources
Medical