. 1982 Jun;56(6):449-52.

Spontaneous breathing with a T-piece circuit: minimum fresh gas/minute volume ratio which prevents rebreathing

  • PMID: 6805364

Spontaneous breathing with a T-piece circuit: minimum fresh gas/minute volume ratio which prevents rebreathing

S E Dean et al. Anesthesiology. 1982 Jun.

Abstract

Thirty adults undergoing elective superficial surgery under enflurane-nitrous oxide anesthesia which intubated and breathing spontaneously via a modified Mapleson D (Brain) T-piece circuit were studied with their consent. Total fresh gas flows which were initially high were adjusted downward until minimal rebreathing (inspired CO2 tension of 5 to 10 mmHg) was present. At this point both fresh gas flow (VF) and minute volume (VE) were recorded, and the ratio of the two (VF/VE) was calculated. The mean VF/VE ratio was found to be 1.89 +/- 0.27 (SD). Linear regression was used to plot VF against VE breathing frequency, tidal volume, age, weight, and end-tidal CO2 tension. Significant correlation was found only with VE (r2 = 0.48, P less than 0.001) and frequency (r2 = 0.44, P less than 0.001). When the ratio VF/VE was plotted against the same variables, no significant correlations was found. This study showed a wide variability in the minimum VF/VE ratio which prevents rebreathing. The respiratory waveform, which was not studied, probably played a role in determining the VF/VE. Nevertheless, 87 per cent of our patients required a VF/VE ratio of 2.0 or less to prevent rebreathing. If one is especially concerned about rebreathing, VE should be measured in the VF adjusted to about twice the measured.

PubMed Disclaimer

Similar articles

Cited by

  • Midazolam and anaesthesia.
    Gemperle M, Kapp W. Gemperle M, et al. Br J Clin Pharmacol. 1983;16 Suppl 1(Suppl 1):187S-190S. doi: 10.1111/j.1365-2125.1983.tb02293.x. Br J Clin Pharmacol. 1983. PMID: 6138076 Free PMC article.
  • Mapleson's Breathing Systems.
    Kaul TK, Mittal G. Kaul TK, et al. Indian J Anaesth. 2013 Sep;57(5):507-15. doi: 10.4103/0019-5049.120148. Indian J Anaesth. 2013. PMID: 24249884 Free PMC article. Review.
  • Anaesthesia breathing circuits.
    McIntyre JW. McIntyre JW. Can Anaesth Soc J. 1986 Jan;33(1):98-105. doi: 10.1007/BF03010918. Can Anaesth Soc J. 1986. PMID: 3081242 Review. No abstract available.

LinkOut - more resources