Spontaneous breathing with a T-piece circuit: minimum fresh gas/minute volume ratio which prevents rebreathing
- PMID: 6805364
Spontaneous breathing with a T-piece circuit: minimum fresh gas/minute volume ratio which prevents rebreathing
Abstract
Thirty adults undergoing elective superficial surgery under enflurane-nitrous oxide anesthesia which intubated and breathing spontaneously via a modified Mapleson D (Brain) T-piece circuit were studied with their consent. Total fresh gas flows which were initially high were adjusted downward until minimal rebreathing (inspired CO2 tension of 5 to 10 mmHg) was present. At this point both fresh gas flow (VF) and minute volume (VE) were recorded, and the ratio of the two (VF/VE) was calculated. The mean VF/VE ratio was found to be 1.89 +/- 0.27 (SD). Linear regression was used to plot VF against VE breathing frequency, tidal volume, age, weight, and end-tidal CO2 tension. Significant correlation was found only with VE (r2 = 0.48, P less than 0.001) and frequency (r2 = 0.44, P less than 0.001). When the ratio VF/VE was plotted against the same variables, no significant correlations was found. This study showed a wide variability in the minimum VF/VE ratio which prevents rebreathing. The respiratory waveform, which was not studied, probably played a role in determining the VF/VE. Nevertheless, 87 per cent of our patients required a VF/VE ratio of 2.0 or less to prevent rebreathing. If one is especially concerned about rebreathing, VE should be measured in the VF adjusted to about twice the measured.
Similar articles
-
Rebreathing improves accuracy of ventilatory monitoring.J Clin Monit. 1995 Nov;11(6):354-7. doi: 10.1007/BF01616740. J Clin Monit. 1995. PMID: 8576717
-
Respiratory waveform and rebreathing in T-piece circuits: a comparison of enflurane and halothane waveforms.Anesthesiology. 1980 Nov;53(5):371-8. doi: 10.1097/00000542-198011000-00003. Anesthesiology. 1980. PMID: 6775562
-
Fresh gas flow in coaxial Mapleson A and D circuits during spontaneous breathing.Acta Anaesthesiol Scand. 1986 Oct;30(7):588-93. doi: 10.1111/j.1399-6576.1986.tb02481.x. Acta Anaesthesiol Scand. 1986. PMID: 3101384
-
Setting the frequency-tidal volume pattern.Respir Care. 2002 Mar;47(3):266-74; discussion 274-8. Respir Care. 2002. PMID: 11874606 Review.
-
[Anesthesia systems].Anaesthesist. 1987 Aug;36(8):393-9. Anaesthesist. 1987. PMID: 3310726 Review. German.
Cited by
-
Midazolam and anaesthesia.Br J Clin Pharmacol. 1983;16 Suppl 1(Suppl 1):187S-190S. doi: 10.1111/j.1365-2125.1983.tb02293.x. Br J Clin Pharmacol. 1983. PMID: 6138076 Free PMC article.
-
Mapleson's Breathing Systems.Indian J Anaesth. 2013 Sep;57(5):507-15. doi: 10.4103/0019-5049.120148. Indian J Anaesth. 2013. PMID: 24249884 Free PMC article. Review.
-
Anaesthesia breathing circuits.Can Anaesth Soc J. 1986 Jan;33(1):98-105. doi: 10.1007/BF03010918. Can Anaesth Soc J. 1986. PMID: 3081242 Review. No abstract available.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources