Abstract
The accuracy and working condition of 210 sphygmomanometers were tested: 100 (50 and mercury and 50 aneroid) models were used in family practices and 100 mercury models in hospitals. Faults in the inflation-deflation system were common and caused mainly by dirt or wear in the control valves. Leakage occurred in 48% of the hospital and 33% of the family practice sphygmomanometers. In the mercury models the mercury or air vents were often in an unsatisfactory condition or the calibrated glass tube dirty. The accuracy of the gauges was examined at 90 and 150 mm Hg: fewer than 2% of the mercury sphygmomanometers but 30% of the aneroid models had errors greater than +/- 4 mm Hg at either pressure. Over half of the cuffs examined had bladders widths less than the recommended size, and 94% had bladders shorter than the length recommended for use on normal adults. Mercury sphygmomanometers should be bought in preference to aneroid models as they are more accurate, less expensive in the long term, and can be maintained by the owner; they should be checked every six to 12 months depending on usage. Replacement parts should be kept readily available.
Similar articles
-
Accurate, reproducible measurement of blood pressure.CMAJ. 1990 Jul 1;143(1):19-24. CMAJ. 1990. PMID: 2192791 Free PMC article. Review.
-
The results of the campaign for evaluating sphygmomanometers accuracy and their physical conditions.Arq Bras Cardiol. 2000 Jan;74(1):31-8. Arq Bras Cardiol. 2000. PMID: 10935290 English, Portuguese.
-
Hidden errors of aneroid sphygmomanometers.Blood Press Monit. 2002 Dec;7(6):309-12. doi: 10.1097/00126097-200212000-00003. Blood Press Monit. 2002. PMID: 12488650
-
How accurate are sphygmomanometers?J Hum Hypertens. 1998 Apr;12(4):245-8. doi: 10.1038/sj.jhh.1000589. J Hum Hypertens. 1998. PMID: 9607693
-
Alternatives to the mercury sphygmomanometer.J Public Health Policy. 2011 Feb;32(1):107-20. doi: 10.1057/jphp.2010.38. Epub 2010 Nov 25. J Public Health Policy. 2011. PMID: 21109765 Review.
Cited by
-
Italian society of hypertension guidelines for conventional and automated blood pressure measurement in the office, at home and over 24 hours.High Blood Press Cardiovasc Prev. 2008 Oct;15(4):283-310. doi: 10.2165/0151642-200815040-00008. Epub 2013 Jan 22. High Blood Press Cardiovasc Prev. 2008. PMID: 23355131
-
Accurate, reproducible measurement of blood pressure.CMAJ. 1990 Jul 1;143(1):19-24. CMAJ. 1990. PMID: 2192791 Free PMC article. Review.
-
Understanding and managing hypertension in an African sub-district: a multidisciplinary approach.Scand J Public Health Suppl. 2007 Aug;69:52-9. doi: 10.1080/14034950701355411. Scand J Public Health Suppl. 2007. PMID: 17676503 Free PMC article.
-
Principles and techniques of blood pressure measurement.Cardiol Clin. 2010 Nov;28(4):571-86. doi: 10.1016/j.ccl.2010.07.006. Cardiol Clin. 2010. PMID: 20937442 Free PMC article. Review.
-
Blood pressure measurement by family physicians.Can Fam Physician. 1988 Feb;34:275-8. Can Fam Physician. 1988. PMID: 21253044 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources