Resistance to gas flow in the "new" anaesthesia circuits: a comparative study
- PMID: 6809297
- DOI: 10.1007/BF03007532
Resistance to gas flow in the "new" anaesthesia circuits: a comparative study
Abstract
The resistance to gas flow is measured in the Jackson-Rees modification of the Ayre's T-piece, the "Bain" circuit, and the circle circuit using the conventionally employed rubber tubing, or either of two coaxial circle tubing systems. The Ayre's T-piece has the lowest resistance to gas flow, followed by the circle circuit with conventional rubber tubing, then the "Bain" circuit and the coaxial circle circuits. When the resistance to gas flow in the inspiratory limb is measured with a tracheal tube attached to each circuit the resistances increase in magnitude, but the differences among the circuits become less significant.
Similar articles
-
A coaxial circle circuit: comparison with conventional circle and Bain circuit.Can Anaesth Soc J. 1982 Mar;29(2):121-5. doi: 10.1007/BF03007989. Can Anaesth Soc J. 1982. PMID: 7066735
-
Coaxial scavenger for paediatric anaesthesia.Can Anaesth Soc J. 1979 Sep;26(5):367-9. doi: 10.1007/BF03006449. Can Anaesth Soc J. 1979. PMID: 487230
-
Modified interface between pediatric ventilator and Ayre's T-piece for use in neonates.Middle East J Anaesthesiol. 1996 Feb;13(4):395-403. Middle East J Anaesthesiol. 1996. PMID: 8713634 Clinical Trial.
-
An economical substitute for Ayre's T piece.Anesth Analg. 2000 May;90(5):1249. doi: 10.1097/00000539-200005000-00062. Anesth Analg. 2000. PMID: 10781500 No abstract available.
-
The Ayre's T-piece turns 80: A 21st century review.Paediatr Anaesth. 2018 Aug;28(8):694-696. doi: 10.1111/pan.13408. Epub 2018 Jun 7. Paediatr Anaesth. 2018. PMID: 29882229 Review.
Cited by
-
Fundamentals of infant anaesthesia.Can Anaesth Soc J. 1984 May;31(3 Pt 2):S45-51. doi: 10.1007/BF03007035. Can Anaesth Soc J. 1984. PMID: 6722633
-
Anaesthesia breathing circuits.Can Anaesth Soc J. 1986 Jan;33(1):98-105. doi: 10.1007/BF03010918. Can Anaesth Soc J. 1986. PMID: 3081242 Review. No abstract available.
-
Roles of endotracheal tubes and slip joints in respiratory pressure loss: a laboratory study.J Anesth. 2016 Oct;30(5):789-95. doi: 10.1007/s00540-016-2210-5. Epub 2016 Jul 1. J Anesth. 2016. PMID: 27370893
-
Role of tube size and intranasal compression of the nasotracheal tube in respiratory pressure loss during nasotracheal intubation: a laboratory study.BMC Anesthesiol. 2017 Oct 17;17(1):141. doi: 10.1186/s12871-017-0432-1. BMC Anesthesiol. 2017. PMID: 29041911 Free PMC article.