Noninvasive Doppler determination of cardiac output in man. Clinical validation
- PMID: 6821902
- DOI: 10.1161/01.cir.67.3.593
Noninvasive Doppler determination of cardiac output in man. Clinical validation
Abstract
A noninvasive technique for assessing cardiac output (CO) was evaluated by comparing it with thermodilution determinations in patients in the intensive care unit. The new method uses pulsed ultrasound to measure aortic diameter and continuous-wave Doppler ultrasound to obtain aortic blood velocity. An initial study evaluating just the velocity measurement showed that changes of the Doppler index of output (DI) correlated well with those of thermodilution cardiac output (TDCO). Linear regression analysis yielded delta DI = 0.87 delta TDCO + 0.14 (r = 0.83, n = 95). Using a university research instrument these measurements were possible in 54 of 60 patients (90%). A second study using a prototype commercial device incorporated the diameter measurement. Ultrasonic cardiac output (UCO), calculated as the time integral of velocity multiplied by the aortic area, was compared to TDCO. The data, obtained from 45 of 53 patients (85%), are described by the linear regression UCO = 0.95TDCO + 0.38 (r = 0.94, n = 110) over a range of 2-11 l/min. Patients with aortic stenosis, aortic insufficiency or a prosthetic valve have been excluded from the second study due to conditions likely to violate the assumptions upon which the calculation of absolute cardiac output is based. These results indicate that accurate CO can be measured by noninvasive ultrasound in most patients. The technique may be useful for extended CO monitoring in acute care patients and for CO assessment in many other types of patients undergoing diagnostic studies and therapeutic interventions.
Similar articles
-
Pulsed wave Doppler measurement of cardiac output from the right ventricular outflow tract.Anesth Analg. 1996 Sep;83(3):466-71. doi: 10.1097/00000539-199609000-00004. Anesth Analg. 1996. PMID: 8780264
-
Comparison between ultrasonic and thermodilution cardiac output measurements in intensive care patients.Crit Care Med. 1986 Apr;14(4):294-7. doi: 10.1097/00003246-198604000-00010. Crit Care Med. 1986. PMID: 3956218
-
Cardiac output quantification by Doppler echocardiography in intensive care--limitations and validation.Rev Port Cardiol. 2000 Jan;19(1):41-64. Rev Port Cardiol. 2000. PMID: 10731790 English, Portuguese.
-
[Noninvasive measurement of cardiac output by pulsed Doppler echocardiography. Correlation with thermodilution].Arch Mal Coeur Vaiss. 1990 Feb;83(2):237-44. Arch Mal Coeur Vaiss. 1990. PMID: 2106860 Review. French.
-
Cardiac output monitoring: a contemporary assessment and review.Crit Care Med. 2015 Jan;43(1):177-85. doi: 10.1097/CCM.0000000000000608. Crit Care Med. 2015. PMID: 25251758 Review.
Cited by
-
Critical Care Ultrasound in Shock: A Comprehensive Review of Ultrasound Protocol for Hemodynamic Assessment in the Intensive Care Unit.J Clin Med. 2024 Sep 10;13(18):5344. doi: 10.3390/jcm13185344. J Clin Med. 2024. PMID: 39336831 Free PMC article. Review.
-
Critical Care Echocardiography as a Routine Procedure for the Detection and Early Treatment of Cardiac Pathologies.Diagnostics (Basel). 2020 Sep 4;10(9):671. doi: 10.3390/diagnostics10090671. Diagnostics (Basel). 2020. PMID: 32899659 Free PMC article.
-
Measurement of aortic blood velocity in acute myocardial infarction by Doppler ultrasound.Intensive Care Med. 1985;11(2):90-4. doi: 10.1007/BF00254781. Intensive Care Med. 1985. PMID: 3886742
-
Measurement of mitral regurgitation by Doppler echocardiography.Br Heart J. 1985 Oct;54(4):384-91. doi: 10.1136/hrt.54.4.384. Br Heart J. 1985. PMID: 4052279 Free PMC article.
-
Doppler-derived cardiac output in healthy newborn infants in relation to physiological patency of the ductus arteriosus.Pediatr Cardiol. 1988;9(2):79-83. doi: 10.1007/BF02083704. Pediatr Cardiol. 1988. PMID: 3399430
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources