Vacuum aspiration with the IPAS Modified Gynecologic Syringe
- PMID: 6839759
- DOI: 10.1016/0010-7824(83)90056-2
Vacuum aspiration with the IPAS Modified Gynecologic Syringe
Abstract
The IPAS Modified Gynecologic Syringe was compared with the electric pump and standard hand pump in a clinical trial of 1227 women undergoing vacuum aspiration. This modified double-valve syringe was designed to fit larger 8 mm, 10 mm and 12 mm cannulae, which are more suitable for late first trimester procedures. The study was conducted at four centers in which the double-valve syringe was used alternately with the vacuum source normally used at each clinic. Although there were notable inter-center differences with respect to certain outcome variables, results were similar for procedures performed at the same center. The only significant finding was the lower proportion of women requiring D & C to complete uterine evacuation among double-valve syringe patients than among hand pump cases at one center. The data indicate that the double-valve syringe is safe and effective for uterine aspiration in women up to 14 weeks' gestation and may be preferable for use in clinics with limited facilities.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical