Goldmann applanation tonometry in patients with regular corneal astigmatism
- PMID: 6869483
- DOI: 10.1016/0002-9394(83)90459-2
Goldmann applanation tonometry in patients with regular corneal astigmatism
Abstract
By averaging vertical and horizontal applanation readings from the Goldmann tonometer, we developed a simple technique that accurately measures intraocular pressure in patients with regular corneal astigmatism. This procedure eliminates the need for keratometry and oblique alignment of the applanator. When the applanator is oriented normally, with the mires displaced horizontally, intraocular pressure is underestimated for with-the-rule corneal astigmatism and overestimated for against-the-rule corneal astigmatism. The error is approximately 1 mm Hg for every 4 diopters of astigmatism. For oblique axis corneal astigmatism, the error is smaller and approaches zero when the axis is either 45 or 135 degrees.
Similar articles
-
Corneal astigmatism in applanation tonometry.Eye (Lond). 2003 Jul;17(5):617-8. doi: 10.1038/sj.eye.6700417. Eye (Lond). 2003. PMID: 12855970
-
Keratometry using the Goldmann tonometer.Eye (Lond). 1993;7 ( Pt 1):43-6. doi: 10.1038/eye.1993.10. Eye (Lond). 1993. PMID: 8285939 Clinical Trial.
-
Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer.Br J Ophthalmol. 2004 Apr;88(4):518-22. doi: 10.1136/bjo.2003.019331. Br J Ophthalmol. 2004. PMID: 15031169 Free PMC article.
-
New ways to measure intraocular pressure.Curr Opin Ophthalmol. 2008 Mar;19(2):122-6. doi: 10.1097/ICU.0b013e3282f391ae. Curr Opin Ophthalmol. 2008. PMID: 18301285 Review.
-
[The role of pachymetry in routine glaucoma diagnosis].Klin Monbl Augenheilkd. 2006 Feb;223(2):117-30. doi: 10.1055/s-2005-858724. Klin Monbl Augenheilkd. 2006. PMID: 16485223 Review. German.
Cited by
-
Comparison between dynamic contour tonometry and Goldmann applanation tonometry.Korean J Ophthalmol. 2009 Mar;23(1):27-31. doi: 10.3341/kjo.2009.23.1.27. Epub 2009 Mar 9. Korean J Ophthalmol. 2009. PMID: 19337476 Free PMC article.
-
Comparative Evaluation of Classic Mechanical and Digital Goldmann Applanation Tonometers.Diagnostics (Basel). 2025 Jul 18;15(14):1813. doi: 10.3390/diagnostics15141813. Diagnostics (Basel). 2025. PMID: 40722562 Free PMC article.
-
Central corneal thickness in adult Chinese.J Huazhong Univ Sci Technolog Med Sci. 2006;26(1):141-4. doi: 10.1007/BF02828062. J Huazhong Univ Sci Technolog Med Sci. 2006. PMID: 16711030
-
Effect of corneal astigmatism on intraocular pressure measurement using ocular response analyzer and Goldmann applanation tonometer.Graefes Arch Clin Exp Ophthalmol. 2010 Feb;248(2):257-62. doi: 10.1007/s00417-009-1202-7. Epub 2009 Sep 29. Graefes Arch Clin Exp Ophthalmol. 2010. PMID: 19787365 Clinical Trial.
-
Graft failure: III. Glaucoma escalation after penetrating keratoplasty.Int Ophthalmol. 2008 Jun;28(3):191-207. doi: 10.1007/s10792-008-9223-5. Int Ophthalmol. 2008. PMID: 18431550 Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical