Liquid chromatographic determination of piroxicam in serum
- PMID: 6879650
- DOI: 10.1097/00007691-198306000-00017
Liquid chromatographic determination of piroxicam in serum
Abstract
A sensitive method for monitoring serum piroxicam (Feldene) is described. The assay requires 1.0 ml of specimen and involves chloroform extraction from an acidified mixture followed by concentration and injection into a liquid chromatograph. Column effluent is monitored at 330 nm. Retention times of piroxicam and the internal standard (naproxen) are 6.6 and 11.0 min, respectively. The lower limit of detection in serum is 0.5 mg/L. Within-day precision (coefficient of variation, CV) of piroxicam in serum (5-13 mg/L range) varied from 3.6 to 6.3%; between-day CV at concentrations of 5 to 20 mg/L varied from 6.5 to 9.8%. Analytical recovery of piroxicam at 20 mg/L was 88%. Several other drugs were analyzed but none interfered with the assay. Serum concentrations found in 28 patients on a 20-30 mg/day dose ranged from 1.5 to 15.2 mg/L.
Similar articles
-
Liquid-chromatographic determination of zomepirac in serum and plasma.Clin Chem. 1982 Mar;28(3):481-4. Clin Chem. 1982. PMID: 7067091
-
Rapid method for the determination of either piroxicam or tenoxicam in plasma using high-performance liquid chromatography.J Chromatogr. 1984 Oct 12;310(2):455-9. doi: 10.1016/0378-4347(84)80116-4. J Chromatogr. 1984. PMID: 6511864 No abstract available.
-
Liquid chromatographic assay of isoxicam in human plasma and urine.J Chromatogr. 1985 Jan 11;337(1):156-9. doi: 10.1016/0378-4347(85)80023-2. J Chromatogr. 1985. PMID: 3980647 No abstract available.
-
Extractionless high-performance liquid chromatographic method for the simultaneous determination of piroxicam and 5'-hydroxypiroxicam in human plasma and urine.J Chromatogr B Biomed Appl. 1995 Nov 3;673(1):142-6. doi: 10.1016/0378-4347(95)00248-h. J Chromatogr B Biomed Appl. 1995. PMID: 8925068
-
Simultaneous determination of piroxicam, meloxicam and tenoxicam in human plasma by liquid chromatography with tandem mass spectrometry.J Chromatogr B Analyt Technol Biomed Life Sci. 2005 Nov 5;826(1-2):214-9. doi: 10.1016/j.jchromb.2005.08.023. Epub 2005 Oct 5. J Chromatogr B Analyt Technol Biomed Life Sci. 2005. PMID: 16169293
Cited by
-
Single and multiple dose pharmacokinetics of a new NSAID (droxicam) in healthy volunteers.Eur J Drug Metab Pharmacokinet. 1989 Oct-Dec;14(4):303-7. doi: 10.1007/BF03190116. Eur J Drug Metab Pharmacokinet. 1989. PMID: 2633925
MeSH terms
Substances
LinkOut - more resources
Full Text Sources