Methicillin-resistant Staphylococcus aureus susceptibility testing with Abbott MS-2 system
- PMID: 6919413
- PMCID: PMC181967
- DOI: 10.1128/AAC.21.4.676
Methicillin-resistant Staphylococcus aureus susceptibility testing with Abbott MS-2 system
Abstract
The antimicrobial susceptibilities of 100 methicillin=resistant Staphylococcus aureus strains were concurrently determined by the Abbott MS-2 System and by the standard disk diffusion method. Agreement between the two methods was 94% or greater for all of the antibiotics tested except for methicillin and gentamicin. This study indicates that the Abbott MS-2 cannot be relied upon for detection of methicillin resistance in clinical S. aureus isolates.