. 1982 Apr;43(4):787-93.
doi: 10.1128/aem.43.4.787-793.1982.

Comparison of four membrane filter methods for fecal coliform enumeration

Comparison of four membrane filter methods for fecal coliform enumeration

J E Pagel et al. Appl Environ Microbiol. 1982 Apr.

Abstract

Four membrane filter methods fecal coliform enumeration were evaluated and compared in six laboratories based on determination of accuracy, specificity, upper counting limit, and recovery comparability. Recovery accuracy with pure cultures ranged from 89 to 100% for m-FC, mTEC (a procedure developed for thermotolerant Escherichia coli), and m-FC2 methods (the latter incorporating a 2-h, 35 degrees C resuscitation period), but was less than 60% for the MacConkey membrane broth method. These figures dropped by approximately 40 to 55% when the cultures were subjected to temperature (10 degrees C) stress. Close to 800 colonies were verified to determine specificity. False-positive colonies occurred most frequently with the m-FC2 method (18%), whereas false-negative colonies were most common on MacConkey membrane broth (26%). In counting range experiments using a variety of samples, the highest upper counting limit was 130 colonies per filter with the mTEC procedure. Recovery comparisons were based on over 130 samples including raw surface waters, raw sewage, and chlorinated and unchlorinated sewage effluents. In general, recoveries were significantly higher with the m-FC2 and mTEC methods; however, on m-FC2, growth of nontarget background organisms was also higher in most cases. Highest recoveries from chlorinated sewage effluents were obtained by the mTEC method, and the addition of a similar resuscitation period to the m-FC procedure (m-FC2) improved fecal coliform recovery from such samples. The best overall performance for fecal coliform enumeration was obtained with the mTEC method with high recovery and low levels of background colonies, good specificity and accuracy, and a high upper counting limit. This procedure also offers the advantage of enumerating E. coli within 24 h.

PubMed Disclaimer

References

    1. Appl Environ Microbiol. 1981 May;41(5):1152-8 - PubMed
    1. Appl Microbiol. 1975 Jun;29(6):826-33 - PubMed
    1. Appl Environ Microbiol. 1976 Oct;32(4):547-52 - PubMed
    1. Appl Environ Microbiol. 1977 Jul;34(1):42-6 - PubMed
    1. Appl Environ Microbiol. 1981 May;41(5):1276-8 - PubMed

MeSH terms