Noninvasive pulsed Doppler determination of cardiac output in neonates and children
- PMID: 7086622
- DOI: 10.1016/s0022-3476(82)80178-9
Noninvasive pulsed Doppler determination of cardiac output in neonates and children
Abstract
Mean blood flow velocity (VAo) in the ascending aorta was measured noninvasively in 33 children, ages 3 days to 17 years, by pulsed Doppler technique at the time of cardiac catheterization. Measurements were made from a suprasternal approach with a portable, range-gated device. The ascending aortic diameter was determined echocardiographically and aortic cross sectional area calculated (A = pi d2/4 cm2). Aortic flow (QAo) was computed from Doppler recordings: QAo (ml/minute) = VAo (cm/second) X A (cm2) X 60 (second/minute). Values were compared with cardiac outputs determined according to the Fick principle, using measured oxygen consumption calculated oxygen capacity, and oxygen saturation. Subjects with aortic valve abnormalities or left ventricular outflow tract obstruction were excluded from study. Agreement between the two methods was excellent (linear regression r = 0.98, slope = 1.07, y-intercept = -4.5 ml, range 403 to 5.540 ml/minute). The Doppler technique is a quick, noninvasive, and accurate method of determining ascending aortic blood flow in neonates and children.
Similar articles
-
Noninvasive measurement of cardiac output in healthy preterm and term newborn infants.Am J Perinatol. 1984 Jan;1(2):148-51. doi: 10.1055/s-2007-999991. Am J Perinatol. 1984. PMID: 6518060
-
Cardiac output determinations in the newborn. Reproducibility of the pulsed Doppler velocity measurement.J Ultrasound Med. 1988 Jun;7(6):311-5. doi: 10.7863/jum.1988.7.6.311. J Ultrasound Med. 1988. PMID: 2969423
-
Effect of patent ductus arteriosus on left ventricular output in premature infants.J Pediatr. 1983 May;102(5):754-7. doi: 10.1016/s0022-3476(83)80252-2. J Pediatr. 1983. PMID: 6842335
-
Cardiac output measurements in congenital heart disease: validation of a simple, portable Doppler method.J Ultrasound Med. 2001 Apr;20(4):365-70. doi: 10.7863/jum.2001.20.4.365. J Ultrasound Med. 2001. PMID: 11316315
-
[Noninvasive determination of stroke volume with impulse Doppler echocardiography: comparison with the Fick method].Herz. 1986 Oct;11(5):269-76. Herz. 1986. PMID: 3781461 German.
Cited by
-
Doppler determination of cardiac output in infants and children: comparison with simultaneous thermodilution.Pediatr Cardiol. 1987;8(4):241-6. doi: 10.1007/BF02427536. Pediatr Cardiol. 1987. PMID: 3432113
-
Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants.Br Heart J. 1992 Feb;67(2):180-4. doi: 10.1136/hrt.67.2.180. Br Heart J. 1992. PMID: 1540440 Free PMC article. Clinical Trial.
-
Cardiac abnormalities in end stage renal failure and anaemia.Arch Dis Child. 1993 May;68(5):637-43. doi: 10.1136/adc.68.5.637. Arch Dis Child. 1993. PMID: 8323332 Free PMC article.
-
The Ratio Fallacy, with Special Reference to the Cardiac Index.Pediatr Cardiol. 2018 Apr;39(4):805-809. doi: 10.1007/s00246-018-1823-x. Epub 2018 Feb 1. Pediatr Cardiol. 2018. PMID: 29387918
-
Haemodynamic features at presentation in persistent pulmonary hypertension of the newborn and outcome.Arch Dis Child Fetal Neonatal Ed. 1996 Jan;74(1):F26-32. doi: 10.1136/fn.74.1.f26. Arch Dis Child Fetal Neonatal Ed. 1996. PMID: 8653431 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous