The heat balance of small babies nursed in incubators and under radiant warmers
- PMID: 7094851
- DOI: 10.1016/0378-3782(82)90100-1
The heat balance of small babies nursed in incubators and under radiant warmers
Abstract
The heat balance of 12 healthy preterm babies (mean birth weight 1.58 kg, gestation 32 weeks, age 7 days) was studied first in an incubator and then under a radiant warmer during normal nursing. Heat production and heat loss by radiation, convection and evaporation were measured in presumed thermoneutral conditions. Although rectal and mean skin temperatures were normal and the same in both environments, there were important differences. Radiation was the major source of heat loss in the incubator and convective losses were low. Under the radiant warmer convection was the major source of heat loss and heat was gained by radiation. A small rise in metabolic heat production occurred under the radiant warmer. Respiratory water loss was low in both environments. Skin water loss was significantly higher under the radiant warmer. The most immature baby (gestation 28 weeks) could not be kept warm in the incubator despite high air temperature, because the evaporative heat loss from her skin was very high. Her body temperatures were normal under the radiant warmer. It is concluded that both devices provide acceptable thermal environments for most preterm babies but that incubators without humidification may be inadequate for immature babies with a high skin water loss.
Similar articles
-
Infant incubators and radiant warmers.Early Hum Dev. 1983 Oct;8(3-4):351-75. doi: 10.1016/0378-3782(83)90018-x. Early Hum Dev. 1983. PMID: 6641579
-
Effect of heat shielding on convective and evaporative heat losses and on radiant heat transfer in the premature infant.J Pediatr. 1981 Dec;99(6):948-56. doi: 10.1016/s0022-3476(81)80030-3. J Pediatr. 1981. PMID: 7310591
-
Radiant energy and insensible water loss in the premature newborn infant nursed under a radiant warmer.Clin Perinatol. 1982 Oct;9(3):483-503. Clin Perinatol. 1982. PMID: 6761034 Review.
-
Thermal balance in term and preterm newborn infants nursed in an incubator equipped with a radiant heat source.Acta Paediatr. 1997 Apr;86(4):403-9. doi: 10.1111/j.1651-2227.1997.tb09031.x. Acta Paediatr. 1997. PMID: 9174228
-
Thermoregulation: incubators, radiant warmers, artificial skins, and body hoods.Clin Perinatol. 1991 Sep;18(3):403-22. Clin Perinatol. 1991. PMID: 1934849 Review.
Cited by
-
Radiant warmers versus incubators for regulating body temperature in newborn infants.Cochrane Database Syst Rev. 2000;2003(2):CD000435. doi: 10.1002/14651858.CD000435. Cochrane Database Syst Rev. 2000. Update in: Cochrane Database Syst Rev. 2002;(2):CD000435. doi: 10.1002/14651858.CD000435. PMID: 10796363 Free PMC article. Updated.
-
Incubation of very immature infants.Arch Dis Child. 1983 Jul;58(7):504-8. doi: 10.1136/adc.58.7.504. Arch Dis Child. 1983. PMID: 6870330 Free PMC article.
-
Artifacts in the measurement of skin temperature under infant radiant warmers.Ann Biomed Eng. 1985;13(5):443-50. doi: 10.1007/BF02407771. Ann Biomed Eng. 1985. PMID: 4073628
-
Humidification of incubators.Arch Dis Child. 1985 Mar;60(3):219-24. doi: 10.1136/adc.60.3.219. Arch Dis Child. 1985. PMID: 3985653 Free PMC article.
-
Open or closed incubators.Arch Dis Child. 1986 Feb;61(2):108-9. doi: 10.1136/adc.61.2.108. Arch Dis Child. 1986. PMID: 3954434 Free PMC article. No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical