A proposed mechanism for transient increases in arterial pressure and flow during angiographic injections
- PMID: 711394
- DOI: 10.1097/00004424-197805000-00004
A proposed mechanism for transient increases in arterial pressure and flow during angiographic injections
Abstract
In a series of animal experiments, we have confirmed the observation that arterial pressure and flow increase distal to the injection site during power injections through non-obstructing catheters. Our data suggest that the phenomenon is secondary to the transient production of turbulence. Thus, for a given injection rate, catheters with smaller end holes create more fluid velocity, which increases the Reynolds number and causes augmentation of downstream pressure and flow. The addition of side holes decreases the fluid velocity of the injectate and minimizes the hemodynamic effects. The discussion deals with factors contributing to the generation of local turbulence, the magnitude of pressure and flow changes caused by the injections, and the clinical implications of these artifacts.
Similar articles
-
[Haemodynamic effects of intraarterial contrast injections (author's transl)].Rofo. 1981 Aug;135(2):143-51. doi: 10.1055/s-2008-1056849. Rofo. 1981. PMID: 6212304 German.
-
New high-flow "cloud" catheter for safer delivery of contrast material.Radiology. 1989 Nov;173(2):461-4. doi: 10.1148/radiology.173.2.2798879. Radiology. 1989. PMID: 2798879
-
Pressure and Flow Rate Changes During Contrast Injections in Cerebral Angiography: Correlation to Reflux Length.Cardiovasc Eng Technol. 2018 Jun;9(2):226-239. doi: 10.1007/s13239-018-0344-3. Epub 2018 Mar 1. Cardiovasc Eng Technol. 2018. PMID: 29497965
-
A solution to the problem of high-flow jets from miniature angiographic catheters.AJR Am J Roentgenol. 1990 May;154(5):1091-5. doi: 10.2214/ajr.154.5.2108548. AJR Am J Roentgenol. 1990. PMID: 2108548
-
Augmented arterial flow and pressure resulting from selective injections through catheters: clinical implications.Radiology. 1978 Apr;127(1):103-8. doi: 10.1148/127.1.103. Radiology. 1978. PMID: 635170
Cited by
-
Hemodynamic effects of rapid injection into the canine superior mesenteric artery.Dig Dis Sci. 1981 Oct;26(10):905-10. doi: 10.1007/BF01309495. Dig Dis Sci. 1981. PMID: 7285730
-
The mixability of angiographic contrast with arterial blood.Med Phys. 2009 Nov;36(11):5064-78. doi: 10.1118/1.3243079. Med Phys. 2009. PMID: 19994517 Free PMC article.
-
The imaging of myocardial perfusion with 81mKr during coronary arteriography.Eur J Nucl Med. 1982;7(9):405-9. doi: 10.1007/BF00256819. Eur J Nucl Med. 1982. PMID: 7128629
-
Angiographic assessment of the efficacy of flow diverter treatment for cerebral aneurysms.Interv Neuroradiol. 2019 Dec;25(6):655-663. doi: 10.1177/1591019919860829. Epub 2019 Jul 11. Interv Neuroradiol. 2019. PMID: 31296064 Free PMC article.
-
Effect of saline injection mixing on accuracy of conductance lumen sizing of peripheral vessels.PLoS One. 2013 Sep 13;8(9):e74622. doi: 10.1371/journal.pone.0074622. eCollection 2013. PLoS One. 2013. PMID: 24058606 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources