Indirect measurement of blood pressure in neonates and infants utilizing an automatic noninvasive oscillometric monitor
- PMID: 7197478
Indirect measurement of blood pressure in neonates and infants utilizing an automatic noninvasive oscillometric monitor
Abstract
An indirect automatic blood pressure (BP) monitor utilizing the oscillometric principle and a microprocessor is now available for use in infants. This device was evaluated by comparing its systolic BP determinations with simultaneous Doppler systolic BP determinations in neonates and infants undergoing general anesthesia. Five determinations each were made in 125 babies who were grouped according to age. Correlation coefficients of the paired BP determinations were: in 27 premature neonates, r = 0.82; in 24 term neonates, r = 0.93; in 47 infants aged 5 to 13 weeks, r = 0.92; and in 27 infants aged 14 to 26 weeks, r = 0.94. The device was also evaluated by comparing its BP determinations with those of indwelling arterial lines present in 20 additional infants (five determinations each). Correlation coefficients were 0.96 for systolic determinations and 0.94 for diastolic determinations. No morbidity or significant technical problems were noted. It was concluded that the device is an accurate, safe, and easy noninvasive method for monitoring BP in infants.
Similar articles
-
An optical oscillometric method for neonatal and premature infant blood pressure monitoring.Adv Neonatal Care. 2009 Apr;9(2):77-81. doi: 10.1097/ANC.0b013e31819d4db3. Adv Neonatal Care. 2009. PMID: 19363328
-
Random zero sphygmomanometer versus automatic oscillometric blood pressure monitor; is either the instrument of choice?J Hum Hypertens. 1995 Nov;9(11):885-9. J Hum Hypertens. 1995. PMID: 8583467 Clinical Trial.
-
Comparison between oscillometric and invasive blood pressure measurements in critically ill premature infants.Acta Paediatr. 2012 Feb;101(2):132-5. doi: 10.1111/j.1651-2227.2011.02458.x. Epub 2011 Sep 23. Acta Paediatr. 2012. PMID: 21880068
-
Method of Blood Pressure Measurement in Neonates and Infants: A Systematic Review and Analysis.J Pediatr. 2020 Jun;221:23-31.e5. doi: 10.1016/j.jpeds.2020.02.072. J Pediatr. 2020. PMID: 32446487
-
Basic principles of noninvasive blood pressure measurement in infants.Adv Neonatal Care. 2005 Oct;5(5):252-61; quiz 262-4. doi: 10.1016/j.adnc.2005.06.005. Adv Neonatal Care. 2005. PMID: 16202967 Review.
Cited by
-
The Dinamap 1846SX automated blood pressure recorder: comparison with the Hawksley random zero sphygmomanometer under field conditions.J Epidemiol Community Health. 1992 Apr;46(2):164-9. doi: 10.1136/jech.46.2.164. J Epidemiol Community Health. 1992. PMID: 1583434 Free PMC article. Clinical Trial.
-
Continuous wave Doppler velocimetry as an adjunct to cross sectional echocardiography in the diagnosis of critical left heart obstruction in neonates.Br Heart J. 1984 Nov;52(5):552-6. doi: 10.1136/hrt.52.5.552. Br Heart J. 1984. PMID: 6238613 Free PMC article.
-
Current state of noninvasive, continuous monitoring modalities in pediatric anesthesiology.Curr Opin Anaesthesiol. 2020 Dec;33(6):781-787. doi: 10.1097/ACO.0000000000000927. Curr Opin Anaesthesiol. 2020. PMID: 33027076 Free PMC article. Review.
-
Accuracy of four indirect methods of blood pressure measurement, with hemodynamic correlations.J Clin Monit. 1990 Oct;6(4):284-98. doi: 10.1007/BF02842488. J Clin Monit. 1990. PMID: 2230858
-
Comparison of the sphygmooscillographic method with the direct and auscultatory methods of measuring blood pressure.J Clin Monit. 1994 Nov;10(6):373-6. doi: 10.1007/BF01618417. J Clin Monit. 1994. PMID: 7836971
MeSH terms
LinkOut - more resources
Full Text Sources
Medical