Assessment of two noninvasive monitors of arterial oxygenation in anesthetized man
- PMID: 7201256
Assessment of two noninvasive monitors of arterial oxygenation in anesthetized man
Abstract
The Hewlett-Packard 47201A ear oximeter and the Radiometer TCM1 transcutaneous oxygen monitor were evaluated for use as noninvasive monitors of arterial oxygenation during inhalational anesthesia in man. Thirty-four healthy adult volunteers were anesthetized to steady states with halothane, enflurane, or isoflurane, and were studied either before or during surgery. Oxygen levels were varied over ranges that included hypoxemia, by manipulating FIO2. Oximeter estimates of SaO2 values and transcutaneous estimates of PaO2 values were compared with conventional measurements of each. Oximeter readings responded rapidly to changes of inspired oxygen concentration and were acceptably accurate estimates of SaO2, except at lower SaO2 levels (less than 80%) during anesthesia without surgery. Transcutaneous oxygen tension readings responded relatively slowly to changes of FIO2 and were frequently inaccurate reflections of PaO2 values. We consider this oximeter suitable as a monitor of arterial oxygenation during anesthesia, but find the transcutaneous electrode unsatisfactory.
Similar articles
-
Simultaneous comparison of three noninvasive oximeters in healthy volunteers.Med Instrum. 1987 Jun;21(3):183-8. Med Instrum. 1987. PMID: 3614041
-
Continuous in vivo measurement of arterial oxygen saturation by oximetry.Heart Lung. 1977 Nov-Dec;6(6):990-3. Heart Lung. 1977. PMID: 244328
-
Ear oximetry: accuracy and practicability in the assessment of arterial oxygenation.Am Rev Respir Dis. 1976 Jun;113(6):745-9. doi: 10.1164/arrd.1976.113.6.745. Am Rev Respir Dis. 1976. PMID: 937815
-
Noninvasive oxygen monitoring.Crit Care Clin. 1988 Jul;4(3):493-509. Crit Care Clin. 1988. PMID: 3063350 Review.
-
Detection and monitoring of hypoxemia and oxygen therapy.Respir Care. 2000 Jan;45(1):65-80; discussion 80-3. Respir Care. 2000. PMID: 10771783 Review.
Cited by
-
Pulse oximetry: evaluation of accuracy during outpatient general anesthesia for oral surgery.Anesth Prog. 1988 Mar-Apr;35(2):53-60. Anesth Prog. 1988. PMID: 3166346 Free PMC article.
-
An update on pediatric monitoring.J Clin Monit. 1985 Jul;1(3):172-9. doi: 10.1007/BF02832258. J Clin Monit. 1985. PMID: 3938480
-
Pulse oximetry in the accident and emergency department.Arch Emerg Med. 1989 Jun;6(2):137-42. doi: 10.1136/emj.6.2.137. Arch Emerg Med. 1989. PMID: 2742667 Free PMC article.
-
Transcutaneous oxygen tension: a physiological variable for monitoring oxygenation.J Clin Monit. 1985 Apr;1(2):130-4. doi: 10.1007/BF02832201. J Clin Monit. 1985. PMID: 3831253 No abstract available.
-