Evaluation of commercially available diagnostic test kits for rubella
- PMID: 7240800
- DOI: 10.1093/infdis/143.4.578
Evaluation of commercially available diagnostic test kits for rubella
Abstract
Three standard hemagglutination-inhibition (HAI) methods were compared with 11 commercially available diagnostic test kits for determination of immunity and serologic diagnosis of rubella using a panel of 100 sera. The three standard HAI methods involved removal of serum inhibitors with kaolin, heparin-MnCl2, or dextran sulfate-CaCl2. The HAI kaolin (Flow Laboratories, McLean, Virginia) and Rubelisa (Microbiological Associates Bioproducts, Walkersville, Maryland) kits gave the best specificity as judged by the absence of false-positive results. Rubacell (Abbott Laboratories, Chicago, Illinois), Rubindex (Ortho Diagnostics, Raritan, New Jersey), Fiax (International Diagnostic Technology, Santa Clara, California), and Rubesure (Calbiochem-Behring, La Jolla, California) gave the best sensitivity as judged by the absence of false-negative results. The kits with the highest degree of both specificity and sensitivity were HAI kaolin (Flow), HAI heparin-MnCl2 (Flow), Rubacell (Abbott), and Rubindex (Ortho). In paired sera from five patients with clinical rubella, seroconversions were shown by seven of the kits. One of the kits, Cordia R (Cordis Laboratories, Miami, Florida), showed no significant rise in antibody titer with one pair of sera. Antibody titers in the same serum varied widely between the different kits.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical